background image

Warning

Prohibition

· 

The motor handpiece designed as Ni-MH battery(rechargeable). The Ni-Cd battery is possible to use but

  the charging time and usage period can be shorten because the charging current is different.

· 

The motor handpiece spend even small electric power during power off. And genially, rechargeable battery is discharging

  little by little when the product is not in use after charged fully. The product is recommended to be charged before use.

· 

This product is not in sterilization once release by manufacturer. Please certainly sterilize before using.(sterilization by user)

·

 Only Rechargeable battery(Ni-MH battery) is allowed to be used. Do not use manganese and alkaline battery.

  If mishandled, that causes leakage or ruptures.

·

 Do not touch the AC power plug with wet hands. There are shock hazard.

·

 Do not drop any kind of liquid and water on the motor handpiece and adapter. The short circuit may cause

  fire, electric shock or corrosion of the product.

·

 Unauthorized modification and dismantling is prohibited.

·

 Be sure to use designated bur by ISO 1797. 

·

 Do not drop the motor handpiece or adapter. Install the adapter on flat ground and safe place.

·

 If the batteries leak, do not allow the liquid to come into contact with your skin, eyes or clothes.

·

 If the batteries leak from the motor handpiece or there change of the exterior or part discolor of the

  motor handpiece, please stop to use and contact to dealer.

·

 If the product is not in use for long period, please put out the battery. There is possible leakage.

·

 Please change the same battery kinds together to use. Otherwise, it may cause leakage and battery rupture.

·

 Please use the designated battery and charger for the product.

·

 The system damages occurred if the product is used near of electromagnetic interference.

·

 Do not use the product near of the flammable product or nitrous oxide.

·

 The product has circuit which can protect of file damage, but the file can be damaged by strong torque or resistance.

·

 Please do not leave the child alone with the product and keep attention to child, elderly and disabled person

  to be close to the product or not.

·

 Do not touch signal input, signal output or other connectors and the patient simultaneously.

·

 External equipment intended for connection to signal input, signal output or other connectors shall comply

  with relevant IEC Standard. (e.g., IEC60950 for IT equipment and IEC60601-1 or IEC60601-1-1 series for

  medical electrical equipment) If in doubt, contact qualified technician or local representative.

3

Содержание Traus ENDO

Страница 1: ...www saeshin com INSTRUCTION MANUAL ENDO...

Страница 2: ...N PRODUCT SPECIFICATION ELECTROMAGNETIC COMPATIBILITY INFORMATION WARRANTY SYMBOL Table of contents This product is cordless handpiece for root canal treatment This motor has to be used only in hospit...

Страница 3: ...pment Type of protection against electric shock Class II equipment Degree of protection against electric shock Type BF applied part Classification according to the degree of protection against ingress...

Страница 4: ...tion the handpiece correctly Pushing it into the charger forcibly in the incorrect direction may cause damage Take care not to put needle pin into the charging terminal of the battery charger Place th...

Страница 5: ...lace If the batteries leak do not allow the liquid to come into contact with your skin eyes or clothes If the batteries leak from the motor handpiece or there change of the exterior or part discolor o...

Страница 6: ...1 10 1 16 1 20 1 32 1 The free running speed of the angle handpiece shall be in accordance with the manufacturer s instructions at a tolerance of 10 as specified Refer to ISO 14457 Prophy products mus...

Страница 7: ...lied to the motor it may automatically step to prevent overheating In this case leave the motor for a while until it cools down Please check all components are included PRODUCT COMPONENTS PART NAME OF...

Страница 8: ...display on the screen or 2 Battery display Shows 3 steps for charging amount 3 Beep display Beep sound function each operating condition 4 Program display Shows the program number currently selected...

Страница 9: ...ile inwards by matching the home position until the angle handpiece clicks Disconnection Press the button on the head to pull out the file Turn off the power when insert or remove the file Pull out th...

Страница 10: ...eep sounds 4 Plug the motor handpiece into the charger with it facing forward The charging lamp lights up and charging starts 5 Orange light is on during charging and green light is on when fully char...

Страница 11: ...act dealer The temperature of motor handpiece is high The temperature of battery is high The charger does not work The charge lamp does not light The plug of the charger is not inserted into the outle...

Страница 12: ...using a nylon brush after soaking for 20 minutes in Enzymatic Detergent Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent Wash the angle handpiece for 10 minut...

Страница 13: ...n lubrication please hold the angle handpiece firmly because the angle handpiece can go out by the spray pressure Do not insert the oil into the motor handpiece It may cause failure When cleaning hand...

Страница 14: ...2 4V Approx 2 5 hours Approx 9 hours No load PRODUCT SPECIFICATION Model name Input Output Power consumption TRAUS XEW10 AC100 240V 50 60HZ 0 1A DC 4 5V 7 5VA Model name Gear ratio Bur replacement ty...

Страница 15: ...he other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Emissions The TRAUS ENDO is intended for use in the electromagnetic environment specified...

Страница 16: ...t 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for 0 5...

Страница 17: ...FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurement...

Страница 18: ...1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These specifications may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection fro...

Страница 19: ...voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction after repairing the product at the not recommended repair shop Normal wear of the con...

Страница 20: ...580 0918 Tel 82 53 587 2341 Fax 82 53 580 0999 www saeshin com traus saeshin com This product is medical device MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park New...

Отзывы: