INTRODUCTION
This product is medical device.
Make sure to read this user manual before using.
This user manual is to assure proper installation and use.
Pay attention to read this user manual in order to making the best use of this product and
to assure the prolonged lifespan before using.
Specially, pay attention the contents with marks such as , and .
<User>
• Qualified Professional
<Intended use>
• ACL(B)-41EP, ACL(B)-42EP, ACL(B)-45EP, ACL(B)-46EP are for indicated for use by dentists in standard
endodontic procedures using rotary files and rotary endodontic drills(Gates-Glidden)
• TRAUS TRP12NN is prophy head connecting DPA Burs.
<Classification of Equipment>
• Type of protection against electric shock: Class II equipment
• Degree of protection against electric shock: Type BF applied part
• Classification according to the degree of protection against ingress of water as detailed
in the current edition of IEC 60529.
- Motor handpiece: IPX0
- Charger: IPX0
•
Not suitable product in the atmosphere that exist flammable anesthetic mixture with air or with oxygen.
• User sterile product (non-sterile product at shipment)
• Classification according to mode of operation: Non-continuous operation
• Applied parts: bur
<Use Time>
• Loading Time: max. 3 minutes
• Resting Time: min. 10 minutes
<Operation & Storage Condition>
1
Operation Condition
Temperature (℃)
10~30
Relative Humidity (%)
10~80
Atmospheric Pressure (hPa)
Temperature (℃)
Relative Humidity (%)
Atmospheric Pressure (hPa)
700~1,060
Transport and Storage
Condition
-20~60
10~90
500~1,060
Содержание Traus ENDO
Страница 1: ...www saeshin com INSTRUCTION MANUAL ENDO...