1. Please read the notice below before operating the product.
1) Make sure that the motor handpiece is fully charged.
2) Make sure to check any contact part to patient.
3) Check the sterile condition of any autoclavable parts.
2. Please read the notice below during the product use.
1) Constantly check the status of the product and patient during surgery.
2) When product or patient has any problem, stop the operation for the patient’s safety and take the proper action.
3) Do not let patient approach or touch the product.
3. Please read the notice below after operating the product.
1) Charge the motor handpiece by plugging it into the charger.
4. When the products are not using for a long term, please be advised to the followings.
1) Remove the battery in the motor handpiece.
2)
Turn off the charger power and unplug the cord. Check the cord regularly and replace it when it is damaged.
3) Clean and sterilize.
4) Keep the products in the recommended condition.
ATTENTION
* If the continuous operation, please put the oil every an hour.
* When lubrication, please hold the angle handpiece firmly because the angle handpiece can go out by the spray pressure.
* Do not insert the oil into the motor handpiece. It may cause failure.
* When cleaning handpiece, do not use solvent systems such as benzene or thinner, but do not wash plastic
products (charger, motor handpiece) with aldehyde.
* Place the angle handpiece in a suitable place or tilt it to allow oil to escape. After the oil is completely dry,
connect to the motor handpiece.
* Be sure to remove any residual oil before fitting the lubricated angle handpiece to the motor handpiece.
* Lubrication spray is not included in the component.
* Caution: Lubrication spray present combustible hazard.
* Use FDA approved lubricant. (K052700 equivalent lubricant is available.)
3. Lubrication
Lubrication product list
●
Angle handpiece, Prophy head
Lubrication method
① Insert the nozzle to motor handpiece assembly part (the end part of angle handpiece/prophy head)
and spray lubricant 2~3 seconds.
② Insert the nozzle tip to head part and spray lubricant 2~3 seconds. (Angle handpiece only)
* Nozzle tip and spray nozzle for angle handpieces are not provided by us.
* Spray nozzle for prophy head are provided by us.
Spray nozzle
Nozzle tip
Spray nozzle