31 

 

Appendix 2_ ECG Analysis Algorithm Accuracy 

 

 This appendix describes the basic function of the Shock Algorithm (SA). 

Overview 

Shock  Algorithm  (SA)  is  an  ECG  analysis  program  in  the  SHOOKA  automatic  external 
defibrillator that recommends whether or not a patient should be given a defibrillation shock. 
This algorithm makes it possible for individuals who are not trained to interpret ECG rhythms 
to provide potentially life-saving therapy to victims of ventricular fibrillation (VF) or pulseless 
ventricular tachycardia (VT). The Shock Algorithm is used to analyze the ECG rhythm during 
the first rhythm analysis after the electrode pads have been placed on the patient when CPR is 
not being performed. It is also used during subsequent rhythm analyses when the user has been 
instructed to stop CPR. 

Automated Interpretation of the ECG 

The SHOOKA automatic external defibrillator recommends a shock if either of the following 
rhythms is detected: 

 

Ventricular fibrillation 

 

Rapid ventricular tachycardia (Rate >150 bpm, patient should be pulseless) 

The SHOOKA automatic external defibrillator recommends no shock for non-shockable ECG 
rhythms as indicated in the Shock Algorithm Performance Report in this appendix. 

The  SHOOKA  automatic external  defibrillator is  designed to detect  and remove pacemaker 
pulses  from  the  ECG  so  that  an  accurate  decision  can  be  reached  while  a  pacemaker  is 
functioning. 

Shock Algorithm 

The Shock Algorithm (SA) in the SHOOKA automatic external defibrillator was verified by 
inputting specific ECG waveform segments from  databases  through the electrode  connector 
and recording the SA decision of 'shock' or 'no shock.' The 'shock' or 'no shock' decision made 
by  the  SA  for  each  ECG  waveform  segment  was  compared  to  the  databases  reference 
annotations or treatment recommendation by similar device. 

The main ECG database used to verify the performance of the SHOOKA automatic external 
defibrillator for  SA is  named the PhysioNet  Test  Set.  In addition,  the ECG database named 
Simulator  Test  Set  was  used  to  provide  shockable  and  Non-shockable  ECG  segments  for 
verification  purposes.  The  following  information  about  the  test  sets  and  the  Summary 

Содержание SHOOKA

Страница 1: ...1 User Manual SHOOKA Electroshock D00983 V7 SAADAT Co...

Страница 2: ...77960719 98 21 77962181 Fax 98 21 77964239 Customer Services Tel 98 21 73098000 98 21 77798910 Cell 98 912 1977157 Fax 98 21 77960761 Legal responsible Address Trionara Technologies AB Polygonv gen 2...

Страница 3: ...f you have any questions about the AED please contact our customer service department This manual is an essential part of the AED system and should always be kept close to it so that it can be obtaine...

Страница 4: ...3 1 BATTERY 12 3 2 PADS 14 4 OPERATION 16 4 1 USE OF THE AED 16 5 AED MANAGEMENT MODE 20 5 1 PREPARING SHOOKA FOR MANAGEMENT MODE 20 5 2 SETTINGS 20 6 MAINTENANCE AND TROUBLESHOOTING 22 6 1 PHYSICAL A...

Страница 5: ...used by trained user with appropriate technical and clinical trainings SHOOKA automated external defibrillator AED are indicated for use on victims of sudden cardiac arrest who are Unconsciousness Abs...

Страница 6: ...of electro surgery Maintenance of the device should be conducted due to the manufacturer recommendations in the user manual Do not connect the AED to a PC or other devices while the device is connecte...

Страница 7: ...nce with CE standard requirements IP22 Under 12 5 mm diameter protection from particles and vertically dripping water have no harmful effect when the enclosure is tilted at an angle up to 15 from its...

Страница 8: ...esponsibilities The manufacturer will not take any responsibility if operator Misuses the device Fails to follow operating instructions Disregards any warning or technical information Modifies the dev...

Страница 9: ...ed by customer requirement 2 Management mode This mode is intended for service and troubleshooting of the device 2 1 Defibrillation SHOOKA uses rectilinear biphasic waveform for defibrillation SHOOKA...

Страница 10: ...In semi automatic versions this button is used to deliver shock to patient after hearing appropriate voice alarm When device is ready to deliver energy to patient the LED indicator beneath this butto...

Страница 11: ...anagement mode touch screen provides required control menu for configuration of the device 6 PAD connector PAD connection 7 Battery placement Battery placement 8 USB Connector USB connection for expor...

Страница 12: ...elow table Table 4 Accessories Manufacturer Part Number Manufacturer Accessory F7958W FIAB Adult PAD F7958P FIAB Child PAD P41021 SAADAT Battery Pack 3 1 Battery SHOOKA uses non chargeable lithium ion...

Страница 13: ...other liquids It may result in fire or explosion To avoid fire and explosion hazard do not burn or incinerate the battery pack Do not use the battery after its expiration date A depleted battery shall...

Страница 14: ...f the pad is not compatible with patient Adult child change the pad with suitable pad according to below steps Adult Connect adult pad as picture below Figure 2 Adult pad placement Child Connect child...

Страница 15: ...d avoid using outdated accessories When using Child pad the Child mode will be selected automatically When using Adult pad for a child patient the rescuer must press the Child button For men with hair...

Страница 16: ...OFF button AED turns ON and applies a self test and visually shows SELF TEST If self test result is true AED visually shows UNIT OK and plays audio prompt as below UNIT OK CALL EMS PRESS CHILD BUTTON...

Страница 17: ...ode or CPR help mode according to previous step 7 Wait until charging is completed Charging is started and AED displays CHARGING x J while charging is completed then AED display this prompt CHARGE COM...

Страница 18: ...rillation air pockets between the skin pads may cause skin burns for the patent Apply pads so that entire electrode adheres to skin Do not reposition the electrodes once applied If the position must b...

Страница 19: ...shock or no shock advised decision Do not touch the patient or the AED during analysis Before using this defibrillator disconnect all equipment that is not defibrillator protected from the patient If...

Страница 20: ...de the device will autosave the data and changes and AED will be turn off 5 1 Preparing SHOOKA for Management Mode 1 Check battery is available 2 Hold ON OFF key for at least 5 seconds 5 2 Settings Th...

Страница 21: ...ted Default state is Semi mode but if user wants to deliver shock automatically he she can select Full mode Factory Settings Settings which should be performed by after sales service or qualified pers...

Страница 22: ...onal accessories are available 6 2 SELF Test SHOOKA provides self test in order to verify device is ready to use or not These tests verify the following features Self test include 1 Battery charge 2 E...

Страница 23: ...tically Visual prompt Replace the AED battery with a new battery pack Device will perform self test again If visual prompt continue contact service department SELF TEST FAILED Visual prompt Turn off t...

Страница 24: ...ars after sale by the company All critical components are tested in this procedure and malfunctioning ones will be replaced Warning Do not open unit remove casing or attempt to repair Dangerous high v...

Страница 25: ...ng and disinfecting Do not sterilize the SHOOKA AED and its accessories LCD Do not wipe the display with dry cloth It might cause scratch Only use a soft sloth with IPA to wipe the polarizer other che...

Страница 26: ...sed Time Battery capacity and Menu ON OFF Button Turns Power ON or OFF Child Selection Button Sets AED to child mode Speaker Provides audio prompts and necessary alarms Shock Button It used for semi a...

Страница 27: ...for all battery conditions Less than 40s Time required for device to get ready at maximum energy from start of ECG analyzing for all battery conditions Less than 30s Compliance and Approvals Classifi...

Страница 28: ...50 J 70 J 85 J 120 J 150 J 200 J 25 43 62 77 94 110 129 50 55 78 97 120 140 172 75 60 85 106 142 166 193 100 58 82 102 140 164 192 125 59 84 94 131 154 180 150 61 87 109 122 143 151 175 57 80 100 113...

Страница 29: ...29 Figure 7 200J Shock Current waveform Figure 8 150J Shock Current waveform Figure 9 120J Shock Current waveform...

Страница 30: ...30 Figure 10 85J Shock Current waveform Figure 11 70J Shock Current waveform Figure 12 50J Shock Current waveform...

Страница 31: ...tricular tachycardia Rate 150 bpm patient should be pulseless The SHOOKA automatic external defibrillator recommends no shock for non shockable ECG rhythms as indicated in the Shock Algorithm Performa...

Страница 32: ...se pacemaker paced rhythm sinus bradycardia supraventricular tachyarrhythmia ventricular escape rhythm ventricular fibrillation ventricular flutter and ventricular tachycardia MIT BIH Arrhythmia Datab...

Страница 33: ...de Intermediate Fine ventricular fibrillation VF 0 20 and 0 1 mV peak to peak amplitude Other VT ventricular tachycardia that does not meet criteria for VT in the shockable rhythms category Also inclu...

Страница 34: ...able NSR Other QRS Asystole 99 Specificity 95 Specificity 95 Specificity 96 97 92 300 total 100 30 100 5270 total 4123 1118 29 Intermediate Fine VF Other VT Report Only Report Only Same as Zoll AED Pr...

Страница 35: ...ge fluctuations flicker emissions IEC 61000 3 3 N A Guidance and manufacturer s declaration electromagnetic immunity The SHOOKA is intended for use in the electromagnetic environment specified below T...

Страница 36: ...e a c mains voltage prior to application of test level Guidance and manufacturer s declaration electromagnetic immunity The SHOOKA is intended for use in the electromagnetic environment specified belo...

Страница 37: ...3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 5500 5785 a For som...

Отзывы: