23
6-3
Troubleshooting
In the following table, issues which could be happen to SHOOKA AED and relative actions
are described:
Table 6: Troubleshooting
Issue
Action
Status indicator shows X
when AED is OFF:
Turn on the AED and wait until automated self-test is
completed. If status indicator still shows X turn off the
AED and replace the battery and repeat the procedure
again. If issue continues, contact service department.
when AED is ON:
Turn the device OFF, after that turn the AED on and wait
until automated self-test is completed. If status indicator
still shows X turn off AED and replace the battery and
repeat the procedure again. If issue continues, contact
service department.
BATTERY IS TOO LOW!!!
Device will shut down
automatically
Visual prompt
Replace the AED battery with a new battery pack.
Device will perform self-test again. If visual prompt
continue, contact service department.
SELF-TEST FAILED
Visual prompt
Turn off the AED and then turn it on and wait until
automated self-test is completed. If status indicator still
shows X , turn off AED and replace the battery and
repeat the procedure again. If issue continues, contact
service department.
INTERNAL ERROR
Voice prompt
Turn off the AED and replace the battery then press
ON/OFF button. Let device performs a self-test. If issue
continues, contact service department.
CHECK DEFIBRILLATOR
PAD
Check defibrillator pad is correctly connected to patient
and AED.
Replace PAD with a new one.
If issue continues, contact service department.
Connect AED Cable to the
Device
Visual prompt
Check defibrillator paddle connector is correctly
connected to the socket.
If issue continues, contact service department.