The Ventilator Setting
Warning
Only trained users should change the ventilator setting.
5-1
Turning on the Ventilator
When the ventilator is turned on,
self test page will appear. The device automatically checks
different mechanical and electronic parts during this test. If there is a failure detect in any
part, “NOT OK” will be shown in red. If even one part except parts which are related to the
memory of the device detected as “Fail”, “FATAL ERROR” wil be shown. In this case after
checking all parts, the menu whould not closed until the problem solve and user turn on and
off the device.
Figure 5-1 FATAL ERROR
The skip key will be activated if it fails in the sections related to checking the memory. The
user can enter the patient page by clicking Skip or shutting down the device with the Shut
down key. It is recommended not to use the device in this situation.