UNIVERSAL TEMPERATURE TRANSMITTER. Rev. Sep. 2012
6
S.T.INNOVATORS
4. TECHNICAL DATA
4.1. Transmitter connection type: two-wire
4.2.
Input:
- Range: adjustable according to customer choice from the whole sensor type
table
- RTD connection diagram:
o
three-wire with auto line resistance compensation
o
two-wire with manual set of line resistance
- Thermocouple:
o
automatic cold junction compensation with external RTD or built in thermo
sensor
o
manual set the temperature for cold junction compensation
4.3. Accuracy:
- For temperature sensors
Cu53, Cu50, Cu100,
type
Е
and type
Т
in whole range
and thermocouples
ПП
1
, type
R
, and type
S,
at temperatures below 200 °
С
-
< 0.5%
- For the rest temperature sensors: 0.2%
±
1LSB.
4.4.
Output current: 4÷20 mA.
4.5. Line resistance: 700
Ω
at 24 VDC.
4.6. Power supply: 18÷36 VDC.
4.7. Dimensions: (W x H x D): 22,5 x 75 x 105 mm.
4.8. Montage: DIN-rail - 35 x 7,5 mm