![RWD D180P Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/rwd/d180p/d180p_user-manual_1503442035.webp)
Autoclaving of
inner
door should be avoided.
Repairing corroded glass is extremely complicated. In most cases, glass
must be replaced if it has been corroded.
5.5 Clean the humidity pan
Clean the humidity pan with soapy water and regular laboratory disinfectants.
Rinse them with sterile water and spray them with 70% alcohol. Humidity pans
can also be sterilized by high temperature.
5.6 HEPA filter maintenance
When the replacement prompt for HEPA filter screen pops up on the message
display, please replace the HEPA filter in time. The user can set a fixed time for
the replacement prompt message of HEPA filter screen. The prompt period can be
set from 0 to 365 days.
5.7 Replace the fuse
1)
Turn off the incubator power switch and unplug the power cord.
2)
Remove the three screws from the back of the top cover with a cross
screwdriver and push back the cover.
3)
Remove the fuse and replace it as shown below
。
4)
Install the top cover by following the dismantling procedure.
Fuse _250V_15A_ broke slowly
Specification
:
0326015.MXP
Fuse-(EFLF1A-5-100P)-5.2*20-1A,
Specification
:
EFLF1A-5-100P