4.9 Alarm and prompt message
Prompts
Reason
Solution
The outer door is
open for too long!
Opening time of outside door >
2min;
1)
Close the outside door
of the equipment.
Please close the inner
door!
Opening time of inside door >
2min;
1)
Close the inside door
of the equipment.
Temperature below
alarm limit!
Actual temperature of inner cavity
< the lower limit of set temperature
alarm threshold;
1)
Adjust the lower alarm
limit;
2)
Check whether the
ambient temperature is
too low;
3)
Contact the after-sales
staff.
Temperature above
alarm limit!
Actual temperature of inner
cavity > the upper limit of set
temperature alarm threshold
1)
Adjust the upper alarm
limit;
2)
Check
whether the
ambient temperature is
too high;
3)
Contact the after-sales
staff.
CO
2
concentration
below alarm limit!
CO
2
concentration < lower limit of
concentration alarm threshold
(Note: if the user sets the CO
2
concentration value to 0.0%, the
alarm can be shielded)
1
)
Check whether the air
pressure valve of
cylinder is abnormal;
2
)
Adjust the preset
pressure of cylinder to
1.0-1.5bar;
3
)
Check whether the air
supply line is abnormal.
CO
2
concentration CO
2
concentration > upper limit of 1
)
Check whether the air