W I R E L E S S   T E L E H E A LT H   G AT E W AY

R T X 3 3 2 0

RTX3320:

The RTX3320 Wireless Telehealth Gateway is used for wireless reception of measurements
taken by on-site patient monitors such as a wireless blood pressure monitor or a wireless
personal weight scale.
When the Gateway has received the measured values, it generates a call and conveys the
data to a remote facility via a public telephone line. In some cases the Gateway is
programmed to deliver data at a specific time of day.
The delivery of data will automatically be repeated if the attempt to deliver data fails. 
The Gateway works in fully automatic mode, and requires no user action.

Cleaning and Maintenance

The Gateway may be cleaned with a soft, dry cloth or a cloth dampened with tap water
and a mild detergent. A household glass cleaner on a soft, dry cloth may be used to clean
off smudges. Never use any solvent, thinner, lighter fluid, alcohol, wet wipes or similar
products on the Gateway.

NOTE

This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. 
This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a par-
ticular installation. If this equipment does cause harmful interference to radio or televi-
sion reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:

Reorient or relocate the receiving antenna.

Increase the separation between the equipment and receiver.

Connect the equipment into an outlet on a circuit different from that to which the 
receiver is connected.

Consult the dealer or an experienced radio/TV technician for help.

This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:

(1) this device may not cause harmful interference, and 
(2) this device must accept any interference received, including interference that may 

cause undesired operation. 

Modifications not expressly approved by this companycould void the user's authority to 
operate the equipment.

This device is for use at home.

Power requirements
Mains:

100-240 VAC, 50-60Hz, 200 mA, Class II, No applied part. 
Only use the original power supply adapter with the type. 
designation FW7333M/06.

Range of temperatures
Operating

10° C to 40°C / 50° to 104° F.  

Storage:

-10° C to +45°C / 14° to 114° F. 

Transport:

-20° C to +60°C / -4° to 140° F.

Relative humidity
Operating:

10% to 90% non-condensing.

Storage/Transport: 10% to 95% non-condensing.

Barometric pressure
Operating:

700 –1060 hPa.

Storage/Transport: 700 –1060 hPa.

Dimensions:

Ø140 x 35 mm / Ø5.5 x 1.4 inches.

Weight:

280 g / 10 oz including power supply adapter.

Radio Transmitter: Bluetooth compliance: Version 1.1. 

Operating frequency: 2.402 to 2.480 GHz.
Output power: 1 to 100 mW / Bandwith: <1MHz.
Antenna type: Internal, omni directional / Antenna gain: -0.8 dBi.
Operational range: Up to 100 meters (indoor) depending of surroundings.  

The frequency band is not harmonized. Use of wireless Bluetooth technology

can be restricted in some countries. Contact local authorities for more information.

The internal antenna used for this mobile transmitter must provide a separation 

distance of at least 20 cm from all persons and must not be co-located or operating in 
conjunction with any other antenna or transmitter.

RTX Healthcare declares that this device is in compliance with the essential

requirements and other relevant provisions of Directive 93/42/ECC. A copy of the
declaration of conformity may be found at the RTX website: www.rtx.dk.

This device has been tested and found to comply with FCC Part 68. Please read the
“US telephone network customer information” in this manual.

This device has been tested and found to comply with TBR21.

This device, including the power supply adapter, is certified to comply with EN60601-1.

This device, including the power supply adapter, is tested according to EN60601-1-2.

Data integrity and security: The privacy of data transmission with this device is secured by
data encryption and authentication (SSL), and meets authorized standards under the HIPAA.

T E C H N I C A L   S P E C I F I C A T I O N S

RTX3320:

The RTX3320 device is for use by patients at home. It is intended to be used in
combination with a variety of patient monitors upon the prescription of a licensed 
physician or other authorized healthcare provider. RTX3320 serves as the remote 
communication link between compatible patient monitors, and the compatible healthcare
facility at another location. The healthcare facility could be at a disease management
center, or with the healthcare/wellness provider or other out-of-hospital caregivers.

The purpose is to transmit selected medical information (such as weight, blood pressure,
blood glucose) over a normal residential telephone line.

RTX3320 is a “black box” device – it is not connected to the patient and not operated by
the patient. The installation must be performed by or under supervision of an authorized
professional healthcare company.

This device does not measure, interpret or make any decisions on the vital data that it
conveys. 

Precautions and warnings

This device is not intended for emergency calls, and may not be used for transmission 
or indication of any real-time alarms or time-critical data.
Clinical judgement and experience are required to check and interpret the measurements
collected and transmitted.

This device is not for use in systems which substitute for medical care.
This device is not intended for patients requiring direct medical supervision or emergency
intervention.

RTX Healthcare · Stroemmen 6 · DK-9400 Noerresundby · Denmark · Tel: +45 9632 2300 · www.rtx.dk

RTX America, Inc. · 2099 Gateway Place · Suite 310  · San Jose · California 95110 · Phone:+1 (408) 441-8600 

M

EDI

EH

U

SE

T WI

EGA

AR

DE

N . 98 5

1 2

4 66 . T

V . 1

00088

U S E R S   M A N U A L

HOW TO INSTALL AND INSTRUCTIONS FOR USE

RTX3320 for PSTN

I N T E N D E D   U S E   O F   R T X 3 3 2 0

IMPORTANT:

Use only the original power supply adapter which comes with the device  

Part# 96100805 US/UK