RS Stock No.712-6039

Instruction Manual

RS-1315

Data Logger Thermometer

FR

EN

IT

DE

ES

TYPE

KJTERSN

MX MN

-

AVG

hour:min

min:sec

INTERVAL

C

K

F

Содержание RS-1315

Страница 1: ...RS Stock No 712 6039 Instruction Manual RS 1315 Data Logger Thermometer FR EN IT DE ES TYPE KJTERSN MX MN AVG hour min min sec INTERVAL C K F ...

Страница 2: ...E OF CONTENTS EN 10 15 10 Version No 00 TABLE OF CONTENTS Title Page 1 INTRODUCTION 1 2 SPECIFICATIONS 2 3 FRONT PANEL DESCRIPTION 4 4 OPERATION INSTRUCTIONS 7 5 MAINTENANCE 13 6 RECALIBRATION PROCEDURE 14 ...

Страница 3: ...s the international temperature scale of 1990 ITS 90 Read the following information carefully before attempting to operate or service the meter When servicing the meter use only specified replacement parts Environment conditions Altitude up to 2000 meters Relatively humidity 80 max Operating ambient temperature 0 to 50 C 32 to 122 F U S Pat No 446 135 Safety symbols Complies with EMC Directive 89 ...

Страница 4: ... R type 2 0 to 1767 0 C 35 to 1999 9 F S type 2 0 to 1767 0 C 35 to 1999 9 F Resolution J K T E and N type 0 1 C F K R and S type 1 0 C F K 0 1 C F K only for reference Measurement accuracy J K T E and N type 0 05 of reading 0 5 C 0 9 F Below 100 C 148 F add 0 15 of reading for J K E and N 0 45 of reading for T R and S type 0 05 of reading 2 C 4 F NOTE This basic accuracy specification does not in...

Страница 5: ...ations Power supply Qty 6 AAA batteries Battery life approx 200 hours carbon zinc battery Auto power off 30 minutes If no key is pressed Low battery indication The BT symbol is displayed when the battery voltage drops below the operating voltage Measurement rate Once per 1 5 seconds Weight 235 gms 8 29 oz Dimensions 150x72x35mm Operating temperature 0 to 50 C 32 to 122 F Operating humidity Below 8...

Страница 6: ...RIPTION EN EN 4 10 15 10 Version No 00 3 FRONT PANEL DESCRIPTION T1 7 6 5 3 4 2 C D 1 B 10 8 9 12 11 A 13 14 TYPE KJTERSN MX MN AVG hour min min sec INTERVAL C K F KJTERSN TYPE hour min min sec INTERVAL MX K MN AVG C F ...

Страница 7: ...ss key to turn the meter on or off 4 MX MN key Press MX MN key to read the maximum minimum and average readings alternately Press MX MN key for 2 seconds to exit MX MN mode 5 MEM key Press MEM key to store a single set of readings in memory Press MEM key for 2 seconds to enter the continuous data logging mode Press again to exit this mode 6 SET key Press SET key to enter to the interval time setti...

Страница 8: ...ey to toggle between the hour min and min sec elapsed time display in MX MN mode 9 TYPE key Press TYPE key to select the thermocouple type K J E T R S or N 10 READ key Press READ key to show manual memory logged readings Press again to exit this mode 11 C F K key Press C F K key to select Celsius C Fahrenheit F or Kelvin K temperature scale 12 key Press backlight key to turn the backlight on and o...

Страница 9: ... circuit the display will show Press C F K key to select the desired temperature scale Press TYPE key to select the required thermocouple type To measure the temperature touch the probe sensor on the object whose temperature is to be measured Read the temperature on the display The display shows OL overload when the temperature being measured is outside the meter valid range 4 2 MAX MIN and AVG Fu...

Страница 10: ...which the MAX MIN or AVG occurred appears on the time display min sec TYPE K MX AVG MN C Press MX MN key for 2 seconds to exit MX MN mode In MX MN mode the C F K and TYPE key are not active 4 3 To erase the memory of datalogger Press key to turn off the meter Press and hold down MEM key then press key to turn on the meter The LCD display will show CLr and all readings in memory will be cleared ...

Страница 11: ... one set of readings will be stored in memory The LCD display shows MEM and a memory location number 01 to 98 Press READ key to enter the manual memory data mode The LCD display will show READ and a memory location number Press or key to scroll through the logged readings Press READ key again to exit READ mode TYPE K C ...

Страница 12: ...for 2 seconds to start logging The LCD display will show MEM and the auto power off function will be disabled The MEM mark will flicker each time a set of data is stored into memory K TYPE C When the memory is full 4200 data sets the FULL symbol will appear on the display and the meter will stop datalogging In the continuous data logging mode the MAX MIN and AVG function can be used Press MEM key ...

Страница 13: ... press key once to enter to alarm high limit value setting mode The LCD display will show ALM Hi mark Press or key until the display shows the desired alarm high limit value and then press key to store the high limit value Repeat to enter the alarm low limit value setting and the LCD display will show ALM Lo mark Press or key until the display shows the desired alarm low limit value and then press...

Страница 14: ...xit the ALARM function press the SET key for 2 seconds then release it 4 7 How to disable Auto Power off Function The meter will automatically turn off if no key press occurs for 30 minutes Press key to turn off the meter Press and hold down the key then press key to turn on the meter and the auto power off function will be disabled The auto power off mark on the time display will not be visible A...

Страница 15: ...ally clean the case with a damp cloth and mild detergent Do not use abrasives or solvents Clean and dry as required 5 2 Battery replacement When the LCD display shows BT the battery has insufficient power to support an accurate test Replace the batteries with Qty 6 type AAA cells ...

Страница 16: ...0 Version No 00 6 RECALIBRATION PROCEDURE The thermometer should be calibrated once a year to ensure its continued accuracy Contact RS Components for further details of calibration service The address is given at the end of these instructions ...