Input Protection Limits
Function
Maximum Input
V DC or V AC
1000 V dc/ac rms
mA AC/DC
800 mA 1000 V fast acting fuse 10 A
A AC/DC
1000 V fast acting fuse (20 A for 30 seconds
max every 15 minutes)
Frequency, Resistance,
Capacitance, Diode Test, Continuity
1000 V dc/ac rms
Temperature
1000 V dc/ac rms
Surge Protection: 8 kV peak per IEC 61010
Heavy Duty True RMS Digital Multimeter / English
Heavy Duty True RMS Digital Multimeter / English
16/12/2017 Version No. 001
16/12/2017 Version No. 001
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of
OVERVOLTAGE CATEGORY I
is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an appropriate low level.
Note-
Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of
OVERVOLTAGE CATEGORY II
is energy-consuming equipment to be
supplied from the fixed installation.
Note-
Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of
OVERVOLTAGE CATEGORY III
is equipment in fixed installations.
Note-
Examples include switches in the fixed installation and some equipment for industrial
use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of
OVERVOLTAGE CATEGORY IV
is for use at the origin of the installation.
Note-
Examples include electricity meters and primar y over-current protection equipment
.
Safety Instructions
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER
apply voltage or current to the meter that exceeds the specified maximum:
2.
USE EXTREME CAUTION
when working with high voltages.
3.
DO NOT
measure voltage if the voltage on the "COM" input jack exceeds 1000 V above
earth ground.
4.
NEVER
connect the meter leads across a voltage source while the function switch is in
the current, resistance, or diode mode. Doing so can damage the meter.
5.
ALWAYS
discharge filter capacitors in power supplies and disconnect the power when
making resistance or diode tests.
6.
ALWAYS
turn off the power and disconnect the test leads before opening the covers
to replace the fuse or batteries.
7.
NEVER
operate the meter unless the back cover, battery and fuse covers are in place
and fastened securely.
8. If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
3. Controls and Jacks
1. 6,000 counts LCD display
2. MAX/MIN button
3. RANGE button
4. Mode button
5. Function switch
6. mA, μA and 10A input jacks
7. COM input jack
8. Positive input jack
9. HOLD and Backlight button
10. Hz and % button
11. RELATIVE button
Note:
Tilt stand and battery compartment
are on the rear of unit.
1
2
6
3
8
9
10
11
4
7
5