13.3 - Appendix - Error list Interrliner
Last updated: 08-10-2018
Error
Extra explanation
Reason/Solution
E2: Communication error! (Board: %s
(%x), Command: %x, TWSR: %x E:
%d)
Communication lost after 3 retries
between Computer and Interrliner.
•
Power cable not connected
on the communication PCB
mounted on the back panel.
•
An I2C cable not connected
•
Serial cable not connected
•
No power on one of the PCB's
•
Short circuit or fault on one
of the PCB's
E3: Measure motor timeout!
Measure head motor did not move or
motor is blocked.
•
Measure head is not at the
Home position.
•
Check the Home sensor.
•
Motor is faulty.
•
Motor driver on drive board is
faulty.
E5: Duplicated ID !!
Sample rejected. Sample already in
carousel.
•
Wait until sample is measured
•
Check general settings (Check
for duplicate ID's)
E6: Program was not properly shut
down. Check settings before
continuing!
There is a possibility that changed
settings which were not saved to disk
are lost.
•
Program stopped and
computer needed to be reset.
•
Computer reset after power
failure.
E7: Needle motor position error!
Timeout! (piercing)
Needle did not go up within a certain
time limit.
•
Needle motor is faulty.
•
Needle motor driver on
needle board is faulty.
•
Needle is blocked.
E8: Fillnozzle not in fill position!
Fill nozzle did not reach the fill position
within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
E9: Air flow failure!
ESR analyser unit was not able to get
a stable reading during the vacuum
test before aspiration the sample.
•
Check for leakage on the
pipette or fill nozzle.
E12: Dilution error: wrong or no
diluent flow. Check the diluter!
Diluter malfunction
•
Check diluent flow sensor
•
Check tubes diluter system
Appendix for Interrliner
126
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...