When the Starrsed RS has been idle for more than eight hours, there might be some backflow of reagents
due to gravity. Prime all tubing before sampling by using the function:
•
Prime all units
All priming functions are sequentially performed one time.
Clean:
•
Wash each pipette:
When the pipette belt turns one position, the pipette at the rinse position will be rinsed and dried,
regardless if it was filled or not.
•
Wash all pipettes:
Before executing this function, check carefully if there are samples in the pipette belt that need to be
measured.
Any remaining samples will be washed away and will
NOT
be measured!
All pipettes on the pipette belt are washed and dried once.
A warning is shown on the display: <Pipette data will be lost!>.
•
Wash only sample pipettes:
Before executing this function, check carefully if there are samples in the pipette belt that need to be
measured. Any remaining samples will be washed away and will NOT be measured!
All pipettes which are currently holding samples are washed and dried once.
A warning is shown on the display: <Pipette data will be lost!>.
•
Fill & Clean:
This button starts the Fill & Clean procedure. During prolonged use of the instrument, proteins are
building up in the Westergren pipettes which need to be removed using a strong cleaning agent. This
function fills all pipettes with a cleaning agent and removes the cleaning agent after a specified time.
•
End-of-day wash:
All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed.
6.5.1.1 - Fill and Clean screen
Fill & Clean:
Automatic fill and clean function, each individual pipette on pipette belt will be filled with cleaning solution.
During prolonged use of the instrument, proteins are building up in the Westergren pipettes which need to
be removed using a strong cleaning agent.
This cycle takes about 90 minutes.
The Fill & Clean function is part of the monthly maintenance procedure.
A warning is shown on the display: <Pipette data will be lost!>.
Starrsed program
68
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...