15.8.4 - One air bubble rising in pipette
1. Usually this is caused by a wet or dirty fill nozzle.
▪ The blood column should not reach right to the base of the pipette.
There must be a clear air gap of 4...5mm at the bottom of each pipette.
2. Insufficient sample volume.
▪ Need more blood in the sample tube.
One air bubble rising can result in ESR Error code 3.
15.8.5 - Small air bubbles rising in pipette
Usually this is caused by a dirty or damaged fill nozzle.
▪
Observe the maintenance schedules.
▪
Clean the fill nozzle.
▪
Check the fill nozzle for damage. If necessary, replace the fill nozzle.
Sample tube is leaking on the fill nozzle side.
▪
Replace the silicon sample tube
Small air bubbles result in ESR Error 3.
Troubleshooting
156
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...