15.4 - Diluter
15.4.1 - Diluter system not sufficiently primed
Before aspirating, the diluent system must be free of air bubbles.
Select
Maintenance-> Prime -> Prime Diluent
, vacuum pump is on and system must be filled with Diluent. When
the diluent is priming, liquid should be seen flowing through the tube connection to the fill nozzle cap.
Occasionally when the diluter system is primed for the first time, air locks occur in the tubing and the diluter
will not self-prime. If this occurs, disconnect the luer fitting at the syringe and connect a syringe filled with
Diluent to the tubing and fill system manually.
1. Check pick-up tube at the Diluent bottle or cubitainer, it may be kinked.
2. One of the Diluent lines has become blocked or kinked.
3. Check if all tubes are still connected.
4. Check if the 2 tubes (ESRI070508) are still connected with the tubes ESRI070406.
5. Check if the 2 tubes positioned correctly in the pinch valve.
15.4.2 - Diluter errors
If the display shows dilution errors it indicates that the current sample has not been diluted correctly e.g.
-21%
Diluent added to the sample. After the sample measurement the dilution rate will be printed as:
EDTA 079.
A (partly) blocked sample flow will be signaled by a time-out on the filling time.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
147
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...