14.2 - Maintenance schedule
Maintenance performed periodically or after number of samples
Maintenance Schedule Interrliner (Example)
Sample volume: 350 per day (5 working
days)
Daily
Weekly
Level 4
(monthly)
Level 3
(quarterly)
Level 2
(halfyearly)
Level 1
(yearly)
Parts
Total
amount/year
or earlier after number of samples
7.500 samples
23.000
samples
46.000
samples
91.000
samples
Per 91.000
samples
Perform End-of-day wash
X
X
X
X
X
X
Clean outside aspiration needle
X
X
X
X
X
X
Check tubing/diluent syringe
X
X
X
X
X
X
Clean outside instrument
X
X
X
X
X
X
Clean Fill nozzle
X
X
X
X
X
Clean waste separator
X
X
X
X
X
Check sensors
X
X
X
X
X
Replace Fill nozzle O-ring
X
X
X
X
QWLV050004
12
Replace bacterial air filter (HEPA)
X
X
X
X
QWLV040002
12
Replace Disc filter (white) Only if internal
waste container is used
X
X
X
X
QWLV040001
12
Replace Rinse tube assembly
X
X
X
X
ESRI090902
12
Replace Saline tube assembly
X
X
X
X
ESRI090903
12
Appendix for Interrliner
104
MRN-164_2-EN Version 2.02 Interrliner