106
HygroClip
communication error
The HygroClip probe does not answer or is not connected
107
calibration error
The difference between the probe reading and the
calibration point is larger than the maximum allowed by the
INI command.
108
calibration error
(overflow)
Internal probe error (or the difference between the probe
reading and the calibration value is too large)
110
unknown reference
probe
The reference probe is not connected or the reference
probe input does not exist
111
Temperature error
During humidity calibration, temperature should be within
the limits of 0 and 80
°
C (32 and 176
°
F)
Warnings:
120
no adjustment
Calibration
121
No HygroClip probe is
connected
Status:
130
the probe was
adjusted
Calibration
9
Test connector for the HygroPalm indicator
NOTE
: The HygroFlex 3 equipped with the optional Ethernet interface does not have a Test
connector.
Definition
The test connector permits to connect a HygroPalm indicator to the HygroFlex (the HygroPalm
goes automatically into its Remote Control Mode). This is especially useful when the transmitter
does not have the optional display and keyboard. Depending on the model of HygroPalm, the
Remote Control mode allows the following:
HygroPalm
1
HygroPalm
2
HygroPalm
3
Display the readings from any probe connected to the
transmitter
x
x x
Display the transmitter status
(configuration, serial number, etc.)
N/A
x x
1-point adjustment of any ROTRONIC HygroClip
digital probe connected to the transmitter, using the
probe connected to the HygroPalm as a reference
N/A
x x
Access the functions
1)
of the transmitter, primarily the
ADJUST M.PT function - see Functions.
N/A N/A
x
1) except for CALCULATE and DISPLAY