The message line of the display shows which probe is the reference: for example REF = Probe 2
indicates that probe #2 is the reference.
Press the ENTER key to accept. After pressing the ENTER key, the message “sure?” should
appear on the LC display. Press ENTER to confirm. Next, the LC display will confirm that the probe
adjustment has been successfully completed. At that time, press ENTER to exit.
Note: you can exit the function at any time (without calibrating the probe) by pressing the MENU
key.
7.6 PROBE
Definition
This function displays the version number and serial number of any ROTRONIC HygroClip digital
probe connected to the instrument.
Procedure
Prior to entering the function, select the probe to be displayed (this is the same as the probe that
was last displayed).
When entering the function, the version number of the probe is displayed first. Use the UP or the
DOWN key to display the serial number of the probe.
7.7 SETTINGS
Definition
This function is used to do the following:
• turn the trend indicators on or off
• change the symbol used for relative humidity
Procedure
Trend
: use the UP or the DOWN key to enable or disable the trend indicators.
The trend of humidity and temperature is shown on the display by an arrow (up or down) to the left
of the measured value. Both arrows are shown to indicate stable conditions. Stable conditions are
defined as rate of change of less than 0.02 %RH / min or °C / min.
Signal stability is first evaluated after 60 seconds into the measurement and is updated every 30
seconds.
HygroFlex 2 and 3: when the trend indicator is enabled, the symbol for the calculated parameter
(e.g. Dp for dew point) is no longer displayed to the left of the numerical value.