Model:
PM-80
Sensor type:
Differential
Operating range:
130mbar
Connector Type:
Push fit connector
Units:
mbar / PSI / kPa / inHg/ mmHg (Hg only will be displayed) / inH20 / mmH20
Operating Temperature:
-10°C to +50°C
Accuracy of Unit Only
:
All ± 1 Digit ( 0.1˚C Low scale; 1˚C
High scale )
From -
10˚C to +50˚C: 0.15%
of r 0.15% of full scale
Negative pressure: 0.3%r 0.3% of full scale
Particular care should be taken not to over pressure the device as this may rupture the sensor membrane. This is
not covered by the manufacturers warranty.
Note: The differential/guage instrument will only measure relative pressures i.e positive pressure applied to the
positive port or negative pressure applied to the negative port.
Instrument Application
The PM-80 is suitable for the measurement of differential pressure.
Instrument Operation
The instrument requires two AA or equivalent cells. These are not supplied with the unit.
When the batteries require changing the following symbol will be displayed:
Please ensure the screws holding the battery compartment are firmly tightened when inserting or replacing batteries
to ensure the IP67 rating is maintained.
Press ON key to turn unit ON. Press OFF key to turn unit OFF.
The unit will switch OFF automatically after 12 minutes, unless any key is activated.
To disable automatic switch off, hold Key when the unit is being switched on. This will disable the automatic
switch-off function until the unit is switched off.
Out of Range
If pressure exceeds the scale range
Out
will appear on the display.
Zeroing
In order to achieve maximum accuracy, it is recommended to zero the instrument in the orientation it will be used
before taking any measurements.
Ensure both ports are open to atmospheric pressure, if indicated pressure is within 3% of full scale of the factory
calibrated zero, press
0.0
key which will adjust the reading to zero.
Hold Function
To enable the Hold
Function, press the “Hold”
button, press again to release the Hold Function.
Hose fitting
The input port(s) on the instrument are designed to accept a hose with 8mm Interior Diameter and 1mm wall. The
hose is fitted by pushing the hose onto the port, an additional clamp fitting may be used to ensure secure fitting.
Smoothing mode
Press
═
key and the instrument continuously averages the last four measurements to give a more stable reading on
the display. Press
═
again to return back to normal settings.
Units of pressure
Press
key to select required units of pressure. Chosen units will be saved when the instrument is switched off.
Back Light
Press and hold
ON
key.