1. Connect the AC adapter with the AC adapter jack in the back
of the unit.
2. Plug the AC adapter into the socket. (AC adapters with re-
quired voltage and current indicated near the AC adapter
jack.)
Caution:
1. Please unload the batteries when operating with
the AC mode for a longer period of time . Leav-
ing the batteries in the compartment for
a long time may cause leakage, which
may lead to damage of the unit.
2. No batteries are needed when
operating with the AC mode.
3. AC adapters are optional. Please contact
the distributor for the compatible AC adapters.
4. Use only the authorized AC Adaptor with this blood pressure
monitor.
12. Using the AC Adapter (Optional)
AC Adapter
(Ø3.8/Ø1.3)
13. Installing Batteries
1. Press down and lift the battery cover in the direction of the arrow to open
the battery compartment.
2. Install or replace 4 “AA” sized batteries in the battery compartment accord-
ing to the indications inside the compartment.
3. Replace the battery cover by clicking in the bottom hooks first, then push in
the top end of the battery cover.
4. Replace the batteries in pairs. Remove batteries when unit is not in use for
extended periods of time.
You need to replace the batteries when
1. low battery icon appears on display.
2. the START/STOP key is pressed and nothing appears on display.
Caution
1. Batteries are hazardous waste. Do not dispose them together with the
household garbage.
2. There are no user serviceable parts inside. Batteries or damage from old
batteries are not covered by warranty.
3. Use exclusively brand batteries. Always replace with new batteries together.
Use batteries of the same brand and same type.
1. Unwrap the arm cuff, leaving the end of the cuff through the D-ring of
the cuff.
2. Put your left arm through the cuff loop. The color strip indication should be
positioned closer to you with the tube pointing in the direction of your arm
(Fig. 1). Turn your left palm upward and place the edge of the arm cuff at
approximately 1.5 to 2.5 cm above the inner side of the elbow joint (Fig. 2).
Tighten the cuff by pulling the end of the cuff.
14. Applying the Cuff
The
National High Blood Pressure Education Program Coordinating
Committee
has developed a blood pressure standard, classifying blood
pressure ranges into 4 stages.
(Ref. The Seventh Report of the Joint Na-
tional Committee on Prevention, Detection,
Evaluation, and Treatment of High Blood
Pressure-Complete Report JNC-7, 2004).
This blood pressure classification are based
on historical data, and may not be directly
applicable to any particular patient. It is
important that you consult with your physi-
cian regularly. Your physician will tell you
your normal blood pressure range as well
as the point at which you will be considered
at risk. For reliable monitoring and refer-
ence of blood pressure, keeping long- term
records is recommended. Please download
the blood pressure log at our website www.
rossmaxhealth.com.
9. Display Explanations
8.
Important Notes: Regarding Your Blood Pressure Measurement
Blood pressure fluctu
-
ates all the time!
You should not be overly
worried if you encoun-
tered two or three meas-
urements at high levels.
Blood pressure changes
over the month and even
throughout the day. It is
also influenced by season
and temperature.
Rossmax HSD (Hemodynamic Stability) technology communicates to a user
if a measurement is taken in the appropriate state of rest ensuring more
correct blood pressure readings. The most common error in blood pressure
measurement is taking the measurement when the users not being at rest
physically or mentally or hemodynamically unstable thus measurements
can be affected and incorrect.
Physical and mental factors can have a significant impact on actual blood
pressure. Simply walking up stairs has a measurable affect on blood pres-
sure. By monitoring changes of the heart rate frequency throughout a
blood pressure measurement, Rossmax HSD can determine if a diagnosis is
taken in the appropriate physiological state of rest. Rossmax HSD prompts
the user to disregard results taken when not at rest and retake another
5. Blood Pressure Standard
6. Blood Pressure Fluctuation
7. Resting Detector Using HSD Technology
The National High Blood Pressure Educa-
tion Program Coordinating Committee
has developed a blood pressure standard,
classifying blood pressure ranges into 4
stages. This unit is equipped with innova-
tive blood pressure risk indicator, which
visually indicates the assumed risk level
(prehypertension / stage 1 hypertension /
stage 2 hypertension) of the result after
each measurement.
This unit is equipped with an Irregular Heartbeat
Detector (IHB) which allows those who have an ir-
regular heartbeat to obtain accurate measurements
alerting the user of the presence of an irregular
heart beat during the measurement.
Note:
It is strongly recommended that you consult
your physician if the IHB icon (
) appears
often.
10. Hypertension Risk Indicator
11. Irregular Heartbeat Detector (IHB)
1.
Date/Time Indicator
2.
Hypertension Risk Indicator
3.
Memory Zones
4.
Resting Detector
5.
Memory Average
6.
Memory Sequence Number
7.
Weak Battery Mark
8.
Irregular Heartbeat Detector (IHB)
9.
Systolic Pressure
10.
Diastolic Pressure
11.
Pulse Rate
12.
Pulse Mark
1. Take your reading in a comfortable environment as measurements can
be affected by hot or cold temperatures. Take your blood pressure at
normal body temperature.
Attention:
Consult the accompanying documents. Please read this manual
carefully before use. For specific information on your own blood
pressure, contact your physician. Please be sure to keep this
manual.
Blood pressure measurements determined with AD761f are equivalent to
those obtained by a trained observer using cuff/stethoscope auscultation
method, within the limits prescribed by the American National Standard, Elec-
tronic or Automated Sphygmomanometers. This unit is to be used by adult
consumers in a home environment. Do not use this device on infants or neo-
nates. AD761f is protected against manufacturing defects by an established
International Warranty Program. For warranty information, you can contact
the manufacturer, Rossmax International Ltd.
1. Introduction
3. Real Fuzzy Measuring Technology
4. Preliminary Remarks
This unit uses the oscillometric method to detect your blood pressure. Be-
fore the cuff starts inflating, the device will establish a baseline cuff pres
-
sure equivalent to the air pressure. This unit will determine the appropriate
inflation level based on pressure oscillations, followed by cuff deflation.
During the deflation, the device will detect the amplitude and slope of
the pressure oscillations and thereby determine for you the systolic blood
pressure, diastolic blood pressure, and pulse.
This Blood Pressure Monitor complies with the European regulations and
bears the CE mark “CE 0366”. The quality of the device has been veri-
fied and conforms to the provisions of the EC council directive 93/42/EEC
dated 14 June 1993 concerning medical devices:
EN 1060-1: 1995/A1: 2002 Non-invasive sphygmomanometers
- Part 1 - General requirements
EN 1060-3: 1997/A2: 2009 Non-invasive sphygmomanometers
- Part 3 - Supplementary requirements for electro-mechanical blood pres-
sure measuring systems
EN 1060-4: 2004 Non-invasive sphygmomanometers
- Part 4: Test Procedures to determine the overall system accuracy of auto
-
mated non-invasive sphygmomanometers.
This blood pressure monitor was designed for long service time. To ensure
continued accuracy, it’s recommended that all digital blood pressure mon-
itors require re-calibration. This monitor (under normal usage with approx.
3 measurements a day) does not require re-calibration for 2 years. Once
the unit should be re-calibrated the device will display
.
The unit should also be re-calibrated if the monitor sustains
damage due to blunt force (such as dropping) or exposure
to fluids and / or extreme hot or cold temperature / humid
-
ity changes. When
appears, simply return to your nearest
dealer for re-calibration service.
www.rossmaxhealth.com
2. Name/Function of Each Part
Date/Time
Set Key
Arm Cuff
LCD Display
User-Switching Key
START/STOP Key
Memory Key
Air Tube and
Connector
Battery Cover
(Located on back of unit)
4“AA”(R06) size,
1.5V batteries.
Data Link
Socket
AC Adapter Jack
Resting
Detector
LED
measurement.
Blood pressure measurement can be affected not only by one’s physiologic
conditions but also physical load, mental stress and certain disorders of
the body:
Physical Causes
Actual Exercise
Post Exercise
Mental Causes
Excitement
Anxiety
White Coat Effect
Distraction
Talking during measurement
Medical Causes
Cardiac Arrhythmia
Impaired Vasal Control
Medication
Hyperventilation
Pulse Period & Pulse Pattern Changes during measurement
Blood Pressure Unstable during measurement
HSD Instability Indication
Hemodynamic Stability
indicates that the circulatory system is sufficiently at rest when taking the
measurement thus measurement results are reliable and good for refer-
ence.
Lack of Hemodynamic Stability
indicates that the circulatory system is not sufficiently at rest (hemody
-
namic instability) when taking the measurement thus the measurement
results should be ignored. Wait a minimum of 5 minutes before re-taking
measurements. More time may be needed depending on one’s physilogic
conditions.
If
icon repeatedly appears, see page 11 for important notes regarding
your blood pressure measurements. If
continues to appear even after
long period of relaxation, this indicates the circulatory system is still not
sufficiently at rest and cannot be relieved within a short period of rest.
Possible causes are, but not limited to, inner unrest of the nervous system,
chronic mental conditions or the developing cardiac arrhythmia that can
cause hemodynamic instability for a
long time and may persist continu-
ously even after long period of rest.
Rossmax HSD is simple to use and
requires no additional user skill
and device adjustments. Rossmax
HSD uses only one measurement to
determine if the users were at rest
(HSD NEGATIVE) or not at rest (HSD
POSITIVE) while taking the measure-
ment.
HSD Negative
(at resting state)
HSD Positive
(not at resting
state)
2
3
7
1
9
10
11
12
8
5
6
4
2. Do not move or talk during measurement as this can elevate readings.
3. Do not move or cross legs during measurements. Keep feel flat on floor
4. It’s suggested that you take your measurements at the same time each
day and use the same arm for consistency.
5. Users should wait a minimum of 5 minutes before taking additional
measurements. More time may be needed depending on your physiol-
ogy.
6. This product is not suitable for people with arryhthmias
7. This device may have difficulty determining the proper blood pressure
for users with diabetes, poor circulation of blood, kidney problems, or
for users who have suffered a stroke.
Instruction
Manual
Model:
AD761 f
EN