![Rosemount Series 8700 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/rosemount/series-8700/series-8700_reference-manual_1487425057.webp)
Reference Manual
00809-0100-4727, Rev EA
October 2010
A-13
Rosemount 8700 Series
Dimensions
See Figure A-9.
Weight
See dimensional tables starting with Table A-9 on page A-21.
ROSEMOUNT 8711 WAFER SENSOR SPECIFICATIONS
8711 FUNCTIONAL
SPECIFICATIONS
Service
Conductive liquids and slurries
Line Sizes
0.15-in. through 8 -in. (4 through 200 mm)
Interchangeability
Rosemount 8711 Sensors are interchangeable with 8732 and 8712E
Transmitters. System accuracy is maintained regardless of line size or
optional features. Each sensor nameplate has a sixteen-digit calibration
number that can be entered into a transmitter through the Local Operator
Interface (LOI) or the Field Communicator. In a digital fieldbus environment,
the 8732E can be configured using any compatible digital fieldbus
configuration tool. No further calibration is necessary.
Upper Range Limit
39 ft./s (12 m/s)
Process Temperature Limits
ETFE Lining
-20 to 300 °F (–29 to 149 °C) for 0.5- through 8-in.
(15–200 mm) line sizes
PTFE Lining
-20 to 350 °F (-29 to 177 °C)
PFA Lining
-20 to 200 °F (-29 to 93 °C)
Ambient Temperature Limits
–30 to 150 °F (–34 to 65 °C)
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
+86-755-8660-6182