Hazardous Area Oxymitter 5000
9-20
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
Figure 9-10. Ceramic Diffuser
Element Replacement
8. Apply a light coating of anti-seize compound to both sides of the new
corrugated seal.
9. Assemble the cell and flange assembly and corrugated seal to the
probe tube. Make sure the calibration tube lines up with the calibration
gas passage in each component. Apply a small amount of anti-seize
compound to the screw threads and use the screws to secure the
assembly. Torque to 4 N·m (35 in-lbs).
10. Apply anti-seize compound to the probe threads, flame arrestor hub,
and setscrews. Reinstall the flame arrestor on the probe. Using pin
spanner wrenches, torque to 14 N·m (10 ft-lbs). Secure the flame
arrestor with the locking setscrew. Torque to 2.8 N·m (25 in-lbs). If
applicable, reinstall the vee deflector, orienting apex toward gas flow.
Secure with the setscrew and anti-seize compound. Torque to 2.8 N·m
(25 in-lbs).
11. On systems equipped with an abrasive shield, install the dust seal
gaskets, with joints 180° apart.
12. If previously removed, install the entire electronics per "Replacement of
Entire Electronics (with Housing)", steps 4 and 5.
13. Follow the instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 5000 into the stack or duct. If there
is an abrasive shield in the stack, make sure the dust seal gaskets are in
place as they enter the 15° reducing cone.
38740023
M5-0.8 X 5 mm
Locking Setscrew
(Use 2.5 mm Hex Key)
M6-1 X 6 mm
Setscrew
(Use 3 mm
Hex Key)
Threaded
Probe Flange
Spanner
Wrench
Ceramic
Diffusion
Element
Vee
Deflector
Cement
Fillet
Cement
Port
Flame
Arrester
Hub
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...