IB-106-340
5-15
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATER T/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
INC
INC
DEC
DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
22220045
Figure 5-13. Fault 13, Invalid Constant
m. Fault 13, Invalid Constant. The CALIBRA-
TION LED flashes twice, pauses for three sec-
onds, and repeats (Figure 5-13).
1. After a calibration has been performed, the
electronics calculates a cell constant value.
2. If the cell constant value is outside of the
range, -4 mV to 10 mV, the alarm will acti-
vate. See paragraph 6-2, Calibration, and
verify the last calibration was performed
correctly.
3. Power down the Oxymitter 4000 and
remove it from the stack.
4. Replace the cell per paragraph 6-8, Cell
Replacement.
Содержание Oxymitter 4000
Страница 19: ...IB 106 340 xv ...