Figure 5-9. Fault 9, High Cell mV
i.
Fault 9, High Cell mV. The O2 CELL flashes
once, pauses for three seconds, and repeats
(Figure 5-9).
1. Using a multimeter, measure across TP1+ to
TP2-.
2. If you measure 1.2 VDC, the cell wires,
either orange or green, have become de-
tached from the input.
3. One possible cause is connector J1. The
orange or green wire has come loose from
the crimped connection.
4. The platinum pad could also be at fault. The
pad could have broken free from the back of
the cell.
5. Replace heater strut per paragraph 6-7,
Heater Strut Replacement. If necessary, re-
place the cell flange assembly per paragraph
6-8, Cell Replacement.