Figure 5-1. Fault 1, Open Thermocouple
a.
Fault 1, Open Thermocouple. The HEATER
T/C LED flashes once, pauses for three seconds,
and repeats (Figure 5-1).
1. Check connector J1. Ensure the connector is
properly seated.
2. Using a multimeter, measure TP3+ to TP4-.
If the reading is 1.2 VDC ±0.1 VDC, the
thermocouple is open.
3. Remove power. Disconnect J1. Measure
continuity across the red and yellow thermo-
couple leads.
4. The measurement should read approxi-
mately 1 ohm.
5. If the thermocouple is open, see
paragraph 6-7, Heater Strut Replacement.