Rosemount 5708 Скачать руководство пользователя страница 1

Quick Start Guide

00825-0200-4570, Rev AA

June 2017

Rosemount

 5708 3D Solids Scanner

Manhole Mounting Tool Kit

Содержание 5708

Страница 1: ...Quick Start Guide 00825 0200 4570 Rev AA June 2017 Rosemount 5708 3D Solids Scanner Manhole Mounting Tool Kit ...

Страница 2: ...rk on the devices must be carried out by manufacturer authorized personnel only Warnings about misuse Inappropriate or incorrect use of the device may result in hazards and application specific malfunctioning such as vessels overfill or damage to system components through incorrect mounting or adjustments If the device is used in a manner not specified in this document the protection provided by t...

Страница 3: ...23 0001 A Installation adapter B Set screw C 2 mm hex key Dimensions are in inches millimeters 1 2 Required equipment Standard tools e g screwdriver pliers 60 mm wrench Wrench to secure the mounting plate bolts 2 mm hex key Two ropes of appropriate type and length see page 6 A B C 1x 3x 1x 4 10 104 Ø 1 73 44 ...

Страница 4: ...June 2017 4 Quick Start Guide 2 0 Install the antenna Step 1 Verify the O rings are in place Step 2 Remove the nut from the neck tube 60 mm ...

Страница 5: ...Quick Start Guide 5 June 2017 Step 3 Mount the installation adapter on the antenna Step 4 Place the mounting plate and nut on the scanner s mounting hole 3x H2 mm Ø 2 05 in 52 mm ...

Страница 6: ...end through the nut mounting plate and mounting hole to the manhole Step 6 Tie the rope to the installation adapter Note Consider securing the antenna by tying an additional rope to the adapter from the manhole opening to prevent the scanner from hitting the side walls Step 7 Carefully lower the antenna into the manhole ...

Страница 7: ...Quick Start Guide 7 June 2017 Step 8 Pull the rope until the neck of the antenna is outside the mounting hole Step 9 Hand tighten the nut ...

Страница 8: ...June 2017 8 Quick Start Guide Step 10 Remove the installation adapter Step 11 Tighten the mounting plate bolts 3x H2 mm ...

Страница 9: ...k Start Guide 9 June 2017 3 0 Rotate the antenna toward vessel center Step 1 Slightly loosen the nut Step 2 Rotate the antenna The notch on top of the thread must be directed toward the vessel center 60 mm ...

Страница 10: ...2017 10 Quick Start Guide Step 3 Tighten the nut Proceed with installation of scanner head See Rosemount 5708 Quick Start Guide and Rosemount 5708 Reference Manual for further installation information 60 mm ...

Страница 11: ...Quick Start Guide 11 June 2017 ...

Страница 12: ...Terms and Conditions of Sale can be found on the Terms and Conditions of Sale page The Emerson logo is a trademark and service mark of Emerson Electric Co Rosemount and Rosemount logotype are trademarks of Emerson All other marks are the property of their respective owners 2017 Emerson All rights reserved Europe Regional Office Emerson Automation Solutions Neuhofstrasse 19a P O Box 1046 CH 6340 Ba...

Отзывы: