5-16
Tank Geometry
5. Choose the
Tank Type
corresponding to the
actual tank. If none of the available options
matches the tank, choose Unknown.
HART command: [2, 1, 2, 1].
Tank Bottom Type
is important for the
measurement performance close to the tank
bottom.
HART command: [2, 1, 2, 2].
Tank Height
is the distance from the Upper
Reference Point to the Lower Reference Point
(see “Tank Geometry” on page 5-3). This
number needs to be as accurate as possible.
HART command: [2, 1, 2, 3].
Select the
Enable Still-Pipe/Bridle
Measurement
check-box and enter the
Pipe
Inner Diameter
if the transmitter is mounted
on a Still-Pipe.
HART command: [2, 1, 2, 4] / [2, 1, 2, 5].
For more information, see “Tank Geometry”
Tank Environment
6. In the Process Condition box, select the
check-boxes that correspond to the conditions
of the tank. Select as few options as possible
and no more than two. See “Process
Conditions” on page 5-5 for more information.
The
Dielectric Chart
lists the dielectric
constants of a large number of products and
can be opened by selecting
View > Dielectric
Constant Chart
menu option.
Choose the
Product Dielectric Range
that
corresponds to the current product. If the
correct range value for this parameter is
unknown, or the contents in the tank are
continually changing, choose
Unknown
.
HART command: [2, 1, 3, 2].
Enter inner
diameter of
the pipe
Содержание 5401
Страница 2: ......
Страница 4: ......
Страница 10: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series TOC 6 ...
Страница 24: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 2 10 ...
Страница 80: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 5 26 ...
Страница 90: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 6 10 ...
Страница 160: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series B 16 ...
Страница 174: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series C 14 ...
Страница 177: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Index 3 Rosemount 5400 Series ...
Страница 178: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series Index 4 ...
Страница 179: ......