Reference Manual
00809-0100-4022, Rev AA
July 2002
A-3
Model 4600
Burst Pressure Limits
Range 4: 11000 psi (758,4 bar)
Temperature Limits
Ambient
–40 to 185 °F (–40 to 85 °C)
Storage
–50 to 230 °F (–46 to 110 °C)
Process Temperature Limits
–40 to 200 °F (–40 to 93 °C)
Turn-On Time
Performance within specifications less than 2.5 seconds after power is
applied to the transmitter
Damping
Analog output response to a step input change is user-selectable from 0.3 to
60 seconds for one time constant. This software damping is in addition to
sensor module response time.
Failure Mode Alarm
HART 4-20mA (output code A)
If self-diagnostics detect a gross transmitter failure, the analog signal will
be driven offscale to alert the user. Rosemount standard and custom
alarm levels are available
(see Table A-1 below).
High or low alarm signal is software-selectable.
Table A-1. Alarm Configuration
High Alarm
Low Alarm
Rosemount
≥
21.75 mA
≤
3.75 mA
Custom levels
(1)
(1) Low alarm must be 0.1 mA less than low saturation and high alarm
must be 0.1 mA greater than high saturation.
20.2 - 23.0 mA
3.6 - 3.8 mA