Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
6-3
NOTE
Spare parts for Rosemount 3152, 3153 and 3154 models are not
hydrostatically tested or nuclear cleaned.
NOTE
The part numbers shown are current at the time of printing of this
manual, but may be revised in the future. Parts provided are
compatible and interchangeable with those listed on your order as to
the form, fit, and function of the part required.
GENERAL CONSIDERATIONS
Because of the nuclear use intended for these parts, certain factors
must be considered regarding maintenance of product qualification and
component traceability during on-site instrument repair. Rosemount
Nuclear Instruments, Inc. rigidly controls the manufacture of each
instrument to ensure that published performance specifications are met
and qualified configurations are maintained. For parts installed outside
of this controlled environment, Rosemount Nuclear Instruments, Inc. is
unable to ensure that the specifications are being satisfied. This
responsibility is shifted to the end user. The integrity of the instrument
as originally assembled is modified.
Replacement of parts has ramifications under 10CFR21, for which the
user is responsible. These same regulations also mandate a component
traceability program, which the user must undertake for the replacement
parts. In view of this, and in order to maintain the qualification of the
product, the user must ensure that all replacement parts are installed in
accordance with the Rosemount Nuclear Instruments, Inc. approved
installation and calibration procedures herein.
SPARE PARTS SHELF LIFE
Store all spare transmitters and spare component parts in accordance
with ANSI N45.2.2 level B.
Qualified transmitters, spare electronic assemblies, spare terminal
blocks, and spare o-rings were qualified based on a shelf life of 20 years
at an ambient temperature of 90°F (32.2°C).
Lubricants and sealant: The date of the end of shelf life (use by date) is
provided with the lubricants and/or sealant, at the time of shipment. The
product has a minimum of six months shelf life at the time of shipment.
All other parts: Shelf life is not applicable.
IMPACT ON TRANSMITTER
QUALIFIED LIFE
Transmitters were qualified based on an installed life of 20 years at an
ambient temperature of 120°F (48.9°C).
The use or installation of spare parts has no effect on overall transmitter
qualified life as established in the baseline qualification.
Содержание 3152
Страница 2: ......
Страница 6: ...Rosemount 3150 Series Reference Manual 00809 0100 4835 Rev BD July 2016 iv ...
Страница 10: ...Rosemount 3150 Series Reference Manual 00809 0100 4835 Rev BD July 2016 1 2 ...
Страница 27: ...Reference Manual 00809 0100 4835 Rev BD July 2016 Rosemount 3150 Series 2 17 ...
Страница 45: ...Reference Manual 00809 0100 4835 Rev BD July 2016 Rosemount 3150 Series 3 18 ...
Страница 51: ...Reference Manual 00809 0100 4835 Rev BD July 2016 Rosemount 3150 Series 4 6 ...
Страница 70: ...Rosemount 3150 Series Reference Manual 00809 0100 4835 Rev BD July 2016 5 19 ...
Страница 77: ...Reference Manual 00809 0100 4835 Rev BD July 2016 Rosemount 3150 Series 6 7 Page Intentionally Left Blank ...
Страница 78: ...Rosemount 3150 Series Reference Manual 00809 0100 4835 Rev BD July 2016 6 8 Page Intentionally Left Blank ...
Страница 79: ...Reference Manual 00809 0100 4835 Rev BD July 2016 Rosemount 3150 Series 6 9 Page Intentionally Left Blank ...