![Rosemount 3144 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/rosemount/3144/3144_manual_1487449010.webp)
Rosemount Model 3144 and 3244MV Smart Temperature Transmitters
2-2
COMMISSIONING: ON THE
BENCH OR IN THE LOOP
The transmitter may be commissioned before or after installation.
However, it may be useful to commission the transmitter on the bench
before installation to ensure proper operation and to familiarize
yourself with its functionality.
Figure 2-1. Installation Flowchart.
START
HERE
Set Units
Set Range
Values
Set Sensor
Type
Set Number
of Wires
Set Damping
Set Jumpers
or Switches
Mount the
Transmitter
Wire the
Transmitter
Power the
Transmitter
Check for
Process Leaks
DONE
Simulate
Sensor Input
No
Yes
No
Yes
Bench
Calibration?
BASIC SETUP
VERIFY
FIELD INSTALL
Within
Specifications?
Refer to