Reference Manual
00809-0100-4825, Rev BB
August 2005
Rosemount 248
3-10
Active Calibrator
The
Active Calibrator Mode
command enables or disables the pulsating
current feature. The transmitter ordinarily operates with pulsating current so
that sensor diagnostic functions, such as open sensor detection and EMF
compensation, can be performed correctly. Some calibration equipment
requires steady current to function properly. By enabling the Active Calibrator
Mode the transmitter stops sending pulsating current to the sensor and
supples a steady current. Disabling the Active Calibrator returns the
transmitter to its normal operating state of sending a pulsating current to the
sensor, thus enabling the sensor diagnostic functions.
The Active Calibrator Mode is volatile and will be automatically disabled when
power is cycled or when a Master Reset is performed using the 375 Field
Communicator.
NOTE
The Active Calibrator Mode must be disabled before returning the transmitter
to the process. This will ensure that the full diagnostic capabilities of the
Rosemount 248 are available.
Disabling or enabling the Active Calibrator Mode will not change any of the
sensor trim values stored in the transmitter.
Sensor Review
The
Signal Condition
command allows viewing or changing the primary
variable lower and upper range values, sensor percent of range, and sensor
damping.
Write Protect
The
Write Protect
command allows you to protect the transmitter
configuration data from accidental or unwarranted changes. To enable the
write protect feature, perform the following procedure:
1.
From the
HOME
screen select
1 Device Setup
,
2 Diag/Service
,
3 Write Protect.
2.
Select
Enable WP.
NOTE
To disable write protect on the Rosemount 248, repeat the procedure,
replacing
Enable WP
with
Disable WP
.
HART Output
The
HART Output
command allows the user to make changes to the
multidrop address, initiate burst mode, or make changes to the burst options.
Fast Key Sequence
1, 2, 2, 1, 3
Fast Key Sequence
1, 4, 2
Fast Key Sequence
1, 2, 3
Fast Key Sequence
1, 3, 3, 3
Содержание 248
Страница 2: ......
Страница 4: ......
Страница 46: ...Reference Manual 00809 0100 4825 Rev BB August 2005 Rosemount 248 4 8 ...
Страница 64: ...Reference Manual 00809 0100 4825 Rev BB August 2005 Rosemount 248 A 18 ...
Страница 74: ...Reference Manual 00809 0100 4825 Rev BB August 2005 Rosemount 248 B 10 ...
Страница 77: ...Reference Manual 00809 0100 4825 Rev BB August 2005 0 1 Rosemount 248 NOTES ...
Страница 78: ...Reference Manual 00809 0100 4825 Rev BB August 2005 Rosemount 248 0 2 NOTES ...
Страница 79: ...Reference Manual 00809 0100 4825 Rev BB August 2005 0 3 Rosemount 248 NOTES ...