UNDERSTANDING SEAL
SYSTEM
PERFORMANCE
Seal Temperature Effects
(Process Temperature
Error)
Fill fluids expand or contract with temperature changes, creating a volume
change that is absorbed by the diaphragm seal and is seen as back pressure at
the transmitter. This back pressure creates a shift in the transmitter reading. For
symmetrical or balanced systems, this error is usually minimal due to the back
pressure being equal on both sides. However, head temperature effect is still
present.
NOTE
Other factors that affect seal temperature effect include diaphragm thickness,
seal type and size, capillary length and inner diameter.
Figure 2-3 on page 2-3 shows how diaphragm size can affect the measurement
reading at the transmitter. For smaller seal sizes, such as the 1
1
/
2
-in. size, the
amount of back pressure on the transmitter causes an additional 12.1 inH
2
O
error. Moving to the 2-in. size gives 1.7 inH
2
O and the largest 3-in. size shown
only has 0.5 inH
2
O error. Using a larger diaphragm can drastically improve
performance and provides a more stable reading.
NOTE
Calculations done in Toolkit with Silicone 200 fill fluid with 3051 Transmitter.
Process Flange
Diaphragm
Flushing Connection
Gasket
Clamp
Содержание 1199
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 8: ...Reference Manual 00809 0100 4002 Rev BB November 2012 Rosemount 1199 TOC 4 ...
Страница 20: ...Reference Manual 00809 0100 4002 Rev BB November 2012 Rosemount 1199 2 10 ...
Страница 38: ...Reference Manual 00809 0100 4002 Rev BB November 2012 Rosemount 1199 3 18 ...
Страница 60: ...Reference Manual 00809 0100 4002 Rev BB November 2012 Rosemount 1199 5 10 ...
Страница 81: ...This page intentionally left blank ...