Rosemount 1159 Скачать руководство пользователя страница 1

Model 1159 Remote 
Diaphragm Seals 
Manual

With Model 1152, Model 1153 Series B, D, and F, and 
Model 1154 Alphaline

®

 Nuclear 

Pressure Transmitters

00809-0100-4504

English

June 1999

Rev. AB

Product Discontinued

Содержание 1159

Страница 1: ...Model 1159 Remote Diaphragm Seals Manual With Model 1152 Model 1153 Series B D and F and Model 1154 Alphaline Nuclear Pressure Transmitters 00809 0100 4504 English June 1999 Rev AB Product Discontinued ...

Страница 2: ...0809 0100 4504 Rev AB June 1999 No Affected Pages Description of Change Effect Date 1 Page i Page ii Back cover page Replace address and phone fax with Rosemount Nuclear Instruments Inc 8200 Market Boulevard Chanhassen MN 55317 USA Tel 952 949 5210 Fax 952 949 5201 10 10 03 ...

Страница 3: ...s product For equipment service needs outside the United States contact the nearest Rosemount representative Within the United States the North American Response Center is at your service 24 hours a day and is a single point contact for all Rosemount equipment service needs If at any time you are not sure what to do have a question about using the product or have a service or support request call ...

Страница 4: ...etween the remote seal and its environment that are essential to meeting IEEE Std 323 1974 requirements To ensure compliance with 10CFR Part 21 the remote seal must comply with the requirements herein throughout its installation operation and maintenance It is incumbent upon the user to ensure that the Rosemount Nuclear Instruments Inc s component traceability program is continued throughout the Q...

Страница 5: ... meter torque specifications to whole numbers in Metal Seal Installation Procedure 3 3 3 4 Rounded newton meter torque specifications to whole numbers in Gasket Installation Procedure 3 3 3 4 Added metric equivalents to dimensions in Mounting of Capillary Clamp 4 1 4 1 Added page reference to step 2 of Gasket Metal Seal Replacement 4 1 4 1 Slightly reworded both notes in Gasket Metal Seal Replacem...

Страница 6: ...iv ...

Страница 7: ...ons 3 1 System Installation Procedure 3 2 Metal Seal or Gasket Installation Procedure 3 2 Mounting of Capillary Clamps 3 4 Maintenance 4 1 Ordering Information 5 1 Model 1152 Ordering Information 5 1 Model 1153 Series B and D Ordering Information 5 2 Model 1153 Series F Ordering Information 5 2 Model 1154 Ordering Information 5 3 Model 1159 Ordering Information 5 3 Model 1159 Spare Parts Ordering ...

Страница 8: ...vi ...

Страница 9: ...ssary to ensure high reliability over an extended service life The remote seal can be supplied with the following pressure transmitters Model 1152 Model 1153 Series B Model 1153 Series D Model 1153 Series F Model 1154 Uses for the remote seal include applications where the environment is unsuitable for instrumentation or where a wet leg reference is required The Model 1159 Remote Seal has been tes...

Страница 10: ...Model 1159 Remote Diaphragm Seals 1 2 ...

Страница 11: ...ength of capillary In addition to temperature effect the length of capillary also affects time response As the length of capillary increases the time response increases Other parameters that must be considered when attempting to define the performance of a remote seal system are vibration effects remote seal stability and static head effects Quality of Fill The presence of air in the fill fluid wi...

Страница 12: ...Model 1159 Remote Diaphragm Seals 2 2 ...

Страница 13: ...ssing on the diaphragm will be read by the transmitter as pressure A misaligned gasket may cause a false reading Remote Seal Installation Considerations Due to infinite mounting configurations mounting of the remote seal is the responsibility of the customer Refer to Rosemount Report D8300151 Qualification Report for Water Filled Remote Seal System for mounting details during qualification testing...

Страница 14: ...t tracing the capillary to ensure a constant ambient temperature For operation at temperatures above 180 F apply a static pressure to the remote seal to prevent the fill fluid from boiling Reference steam tables for appropriate pressures System Installation Procedure Unbolt the seal upper and lower housing they are factory assembled but not torqued and thread the lower housing to the hard piping s...

Страница 15: ... tighten Continue placement of bolts and nuts until all eight are in place and finger tight 7 Tighten the bolts and nuts in a crossing pattern to 5 ft lb 7 N m 8 Repeat to 10 ft lb 14 N m 9 Repeat to 15 ft lb 20 N m 10 Repeat to 17 ft lb 23 N m 11 Repeat to 19 ft lb 26 N m 12 Repeat to 22 ft lb 30 N m 13 Repeat to 27 ft lb 37 N m 14 Recheck at 27 ft lb 37 N m Gasket Installation Procedure NOTE Use...

Страница 16: ... and nuts until all eight are in place and finger tight 7 Tighten the bolts and nuts in a crossing pattern to 10 ft lb 14 N m 8 Repeat to 15 ft lb 20 N m 9 Repeat to 20 ft lb 27 N m 10 Recheck at 20 ft lb 27 N m Mounting of Capillary Clamps Clamps P N 01159 0301 0001 should be placed approximately 18 in 46 cm apart to hold the capillary tubing in place using any standard grade 20 screws The minimu...

Страница 17: ...tal seal installation see page 3 2 after discarding the existing gasket metal seal NOTE The procedure for replacement of the gasket or metal seal must be followed whenever the remote seal is disassembled For servicing other than changing out electronics boards transmitter and remote seal system must be returned to Rosemount NOTE See RETURN OF MATERIAL at the front of this manual or in the transmit...

Страница 18: ...Model 1159 Remote Diaphragm Seals 4 2 ...

Страница 19: ...a 0 170 to 0 1 000 psi 0 1 15 to 0 6 89 MPa Code Input N 1 4 20 mA dc with Adjustable Damping MATERIALS OF CONSTRUCTION Code Flanges Drain Vent Valves Isolating Diaphragms Electronics Housing Covers 22 92 316 SST 316 SST 316 SST 316 SST 316 SST 316 SST Aluminum Austenitic SST Code T Options T1808 2 T1809 One Seal with one NPT V D Valve Two Seals Code Options PB PM Panel Mounting Bracket 2 in Pipe ...

Страница 20: ...mes the responsibility for qualifying the process interface on this option 2 Unless otherwise noted flange options D E and F will have the remote seal installed on the high side MODEL 1153 SERIES F ORDERING INFORMATION Model Description 1153 Alphaline Pressure Transmitters for Nuclear Applications Code Pressure Measurement D Differential Pressure 2 000 psig Static Pressure Rating Code Series F Pai...

Страница 21: ... remote seal installed on the high side MODEL 1159 ORDERING INFORMATION Model Product Description 1159 Remote Seal Assembly Code Threaded Remote Seal A B C D 1 in NPT Process Connection in NPT Process Connection 1 in NPT Process Connection 1 in NPT Socket Weld Connection Code Capillary 2 05 10 15 20 25 30 35 40 45 50 55 60 5 ft 10 ft 15 ft 20 ft 25 ft 30 ft 35 ft 40 ft 45 ft 50 ft 55 ft 60 ft Code...

Страница 22: ...N 1 Part Description Part Number Metal Seal Capillary Clamp Lower Housing 18 NPT 14 NPT 1 11 NPT Nut Bolt Kit 8 each 01159 0300 0002 01159 0301 0001 01153 0301 0001 01153 0301 0002 01153 0301 0003 01159 0320 0001 1 Refer to transmitter instruction manual for transmitter spare parts ...

Страница 23: ......

Страница 24: ...clear Instruments Inc 12001 Technology Drive Eden Prairie MN 55344 Tel 612 828 8252 Telex 4310012 Fax 612 828 8280 1999 Rosemount Nuclear Instruments Inc http www rosemount com 00809 0100 4504 Rev AB 00809 0100 4504 ...