background image

Fingertip 

                                     

Pulse Oximeter

 

USER MANUAL 

Ver2.0C2 

General Description 

Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin 
(Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the 
Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. 
Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, 
Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, 
asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems 
in a timely manner. 

The fingertip pulse oximeter features low power consumption, convenient operation and portability. Place one fingertip into the 
photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. It has been proven in clinical 
experiments that it also features high precision and repeatability. 

Measurement Principle 

Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to 
Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. 
Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity 
Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near 
infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a 
photosensitive element, will be shown on the oximeter’s display through process in electronic circuits and microprocessor shown on 
the oximeter’s display through electronic circuits and a microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Red and Infrared-ray Receipt Tube 

Precautions For Use 

Before use, carefully read the manual. 

Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU). 

The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO

2

 measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO

2

 measurement. 

Do not use the fingertip pulse oximeter in an MRI or CT environment. 

Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms. It is not for 
continuous monitoring. 

Do not use the fingertip pulse oximeter in an explosive atmosphere. 

The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other 
methods of assessing clinical signs and symptoms. 

Check the pulse oximeter sensor application site every 4 hours to determine the positioning of the sensor and circulation 
and skin sensitivity of the patient. 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 
intended for sterilization. 

10 

Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, 
including batteries. 

11 

This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or 
systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical 
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or 
strength of a source might disrupt the performance of this device. 

12 

Portable and mobile RF communications equipment can affect medical electrical equipment. 

Rx only: “Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.” 

Inaccurate measurements may be caused by 

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin); 

Intravascular dyes such as indocyanine green or methylene blue; 

High ambient light. Shield the sensor area if necessary; 

Excessive patient movement; 

High-frequency electrosurgical interference and defibrillators; 

Venous pulsations; 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line; 

The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia; 

The patient is in cardiac arrest or is in shock; 

10 

Fingernail polish or false fingernails; 

11 

Weak pulse quality (low perfusion); 

12 

Low hemoglobin; 

Product Properties 

Operation of the product is simple and convenient. 

The product is small in volume, light in weight and convenient to carry. 

Power consumption of the product is low and the two AAA batteries can be operated continuously for 30 hours.   

A low voltage warning will be indicated when battery voltage is low and normal operation of the oximeter might be influenced. 

The product will automatically power off when there is no signal for longer than 8 seconds. 

Intended Use 

Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin 
(SpO

2

) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, 

intensive care etc). It is not for continuous monitoring. 

The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement of batteries. 

Operation Instructions 

Install two AAA batteries according to the Battery Installation instructions listed above in the right column. 

Open the clamp as illustrated in the picture below. 

Fully insert one fingertip into the silicone hole of the oximeter before releasing the clamp. 

Press the switch button once on front panel. 

Keep your finger still during measurement. 

Read corresponding data from display screen. 

Press the button again to toggle between six display modes. 

After turning on the Oximeter, each time you press the power switch, the Oximeter will switch to another display mode. 
There are 6 display modes shown as follows: 

1. 

          2. 

          3. 

 

4. 

          5. 

          6. 

 

Holding the power switch for longer than one second, will adjust the brightness of the oximeter. There are 10 levels of brightness. 
The default level is level four. 

Front Panel 

           

 

Patient pulse quality signals are indicated by bar graph. The bar is graded as 10 levels, if the strength is level 2 to 3, the pulse signal 
is inadequate. 

Product Accessories 

1. One 

lanyard 

2. Two 

AAA 

batteries 

3. 

One instruction manual

 

Battery Installation 

1. 

Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) 
signs in the compartment. If the polarities are not matched, damage may be caused to 
the oximeter. 

2. 

Slide the battery door cover horizontally along the arrow shown as the picture. 

Notes:  

—

 

Install the batteries with the correct polarity. Incorrect placement may cause damage to 
the bracket. 

—

 

Please remove the batteries if the pulse oximeter will not be used for long periods of time. 

Using the Lanyard 

1. 

Thread thinner end of the lanyard through the hanging hole. 

2. 

Thread thicker end of the lanyard through the threaded end before pulling it 
tightly. 

 

Warnings! 

1. 

Keep the oximeter away from young children. Small items such as the battery door, 

battery, and lanyard are choking hazards. 

2. 

Do not hang the lanyard from the device’s electrical wire. 

Maintenance and Storage 

1. 

Replace the batteries in a timely manner when low voltage lamp is lighted. 

2. 

Clean surface of the fingertip oximeter before it is used in diagnosis for patients. 

3. 

Remove the batteries if the oximeter is not operated for a long time. 

4. 

It is best to store the product in -20

℃~+

55

 

and 

93% humidity

.

 

5. 

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage. 

6. 

Dispose of battery properly; follow any applicable local battery disposal laws. 

Cleaning the fingertip pulse oximeter 

Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70% 
isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test. 

Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to 
dry thoroughly before reuse. 

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to establish 
the SpO

2

 accuracy. The measured arterial hemoglobin saturation value (

SpO

2

) of the sensors is compared to arterial hemoglobin 

oxygen (

SaO

2

) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to 

the CO-oximeter samples measured over the 

SpO

2

 range of 70 

 100%. Accuracy data is calculated using the 

root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment

Particular requirements for the 

basic safety and essential performance of pulse oximeter equipment for medical use. 

Specifications 

1. Display Type 

OLED display 
Data update time: <2s 

2. SpO

2

 

Display range: 0-99% 
Measurement range: 70-99%   
Accuracy: 70%-99%: ±3%;    0%~69% no definition 
Resolution: 1%   

3. Pulse Rate 

Display range: 0~254BPM 
Measure range: 30-235 BPM 
Accuracy: 30~99bpm, ±2bpm;    100~235bpm, ±2% 
Resolution: 1BPM 

4. Probe LED Specifications 

 

Wavelength Radiant 

Power 

RED 660

±

2nm 1.8mW 

IR 940

±

10nm 2.0mW 

5. Power Requirements 

Two AAA alkaline Batteries 
Power consumption: Less than 40mA   
Battery Life: Two AAA 1.5V, 600mAh alkaline batteries could be continuously operated as long as 30 hours.   

It is equipped with a function switch, through which the oximeter can be powered off in case no finger is the oximeter longer than 8 
seconds. 

6. Outline Dimension 

Length: 58mm 
Width: 32mm 
Height: 34mm 
Weight: 50g (including two AAA batteries) 

7. Environment Requirements 

Operation Temperature: 5

40

 

Storage Temperature: -20

~+

55

 

Ambient Humidity: 15%~80% in operation 

93% in storage 

8. Equipment Response Time 

As shown in the right figure. 

RMI-POX4C 

Отзывы: