Maintenance
Notification!
Failing to comply with periodic maintenance may result of degrading of product function and
safety.
Notification!
Ropox will make parts list and circuit diagrams available for maintenance and repair beyond
what is described in this manual. Repair beyond what is described in this manual may only be
done by Ropox designated service personnel.
6.1 Periodic maintenance of bed guards
If the product is used in a humid or dry atmosphere the locking pin may be prone to failure from
lubricants drying. Besides checking the function of the locking mechanism before every use, a periodic
lubrication of the locking pin every 6 month is recommended.
6.2 Periodic charging of battery
The battery must be periodically charged minimum every 6 months to sustain capacity. If the product is
left unused or the battery is disconnected this may result degrading of the battery capacity. Charging the
battery minimum every 6 months will mitigate this.
6.3 Replacement of battery
Notification!
The battery is to be replaced after 4 years at the latest.
6.4 Fault finding
6.5 Calibration of lifting columns
Lifting columns are aligned and calibrated from the factory. In case they need calibration to run
simultaneously use the
“Reset”
procedure.
Problem
Check the following
Mobilio does not move up or down when the hand-
control switch is pressed
1. The product is connected to 230V and the
control box LED indicator lights green
2. Motor cables are connected to control box
3. The control switch has been connected with
the control unit
4. If the product is used with battery, check
whether it is charged.
5. The motors are overloaded (note: 2 min.
ON, 18 min OFF).
Control switch defective
1. Order a new control switch
Mobilio is not horizontal / level
”Reset”
Press and hold UP ( ) and DOWN ( ) buttons.
Keep the buttons depressed until the control unit
has beeped approx. ten times. After that, the
product may be adjusted to the bottom position for