Rooster™ Monitor200 USER MANUAL

 

 

 

 

 

  

Rooster™ Monitor200 USER MANUAL 62330MN000

-A01

 

    

 1 of 51 

 

User Manual 

 

 

High Performance, BACnet® MS/TP Capable Airflow Monitor 
with 
Full-Color Touch Screen Display 
Version 1.06 
 

Complete Kit includes: 

o

 

Alarm Module 

o

 

Choice of Sensor Type 

o

 

Power Supply 

 

 

 

 

 

 

This is proprietary information of Degree Controls Inc., contents are protected under US copyright laws ©Degree Controls, Inc. 2021. 

Содержание Monitor200

Страница 1: ...Performance BACnet MS TP Capable Airflow Monitor with Full Color Touch Screen Display Version 1 06 Complete Kit includes o Alarm Module o Choice of Sensor Type o Power Supply This is proprietary infor...

Страница 2: ...edure 16 Probe Sensor 17 Sidewall Sensor 18 Inline Sensor 22 Boot Up Procedure 23 Access Tier Privileges USER EH S CERTIFIER 23 Set Change Passwords 24 Calibration 24 Calibration Airflow Sensor 24 Tro...

Страница 3: ...ER MANUAL 62330MN000 A01 3 of 51 Links to Relay Outputs 45 I O Polarity 46 Import Export 46 EEPROM 46 Event Log 46 USB Field Upgrade Procedure 47 Soft Reboot from Info Screen 48 Power Cycle Procedure...

Страница 4: ...oosterTM Monitor200 s many features including but not limited to Glove friendly touchscreen Best in class air velocity sensing Choice of sensor for canopies negative pressure cabinets ducted flows or...

Страница 5: ...e In all other cases this LED will remain off Soft Buttons and Home Screen Layout Feature Overview CUSTOM ASSET TAG Users logged in as EH S can change default Asset Tag to a custom tag This is useful...

Страница 6: ...System Alarms Sound Device Volume ON SCREEN ALERT BANNERS A yellow banner will appear beneath on screen velocity readings to alert users regarding various scenarios such as low airflow or power failu...

Страница 7: ...must be closed See the Alarm Threshold Configuration section for instructions on how to setup night setback alarm thresholds Home Settings System Alarms Sound Alarm Controls Night Setback ALARM LATCH...

Страница 8: ...e RoosterTM Monitor200 uses a two point calibration procedure to collect both a high velocity and low velocity reference point Note For a probe sensor calibration the low velocity reference point is c...

Страница 9: ...t Number Phoenix Contact 1803659 2 position orderable Part Number Phoenix Contact 1803578 Table 1 Connections Rear Panel of Display Module Optional Sash Switch Kit Degree Controls Part 62310AS004 incl...

Страница 10: ...nt using the orderable mating connectors listed above Sash Input The RoosterTM Monitor200 provides you with the ability to select one of two sash input settings 1 Digital sash sensor or 2 Analog sash...

Страница 11: ...3 on the input connection block Analog Sash With the sash input configured as Analog the Monitor200 monitors the sash position as detected from the 0 10V analog signal on the analog input The design i...

Страница 12: ...threshold Note NSB threshold is independently set Muted In this state when the night setback signal is asserted the Monitor200 will alarm screen will display a red background and the LED will flash wh...

Страница 13: ...fan or damper Once Control Output is enabled the user needs to enter or select 1 Setpoint desired face velocity or volumetric flow if volumetric flow is enabled and 2 Output a 0 20 mA b 4 20 mA c 0 5...

Страница 14: ...A c 0 5 V d 0 10 V 2 Max Scale upper limit of scale For example with Output set to 0 5 V Max Scale set to 10 m s and flow in the cabinet set such that 5 m s is displayed on the RoosterTM the analog ou...

Страница 15: ...Rooster Monitor200 USER MANUAL Rooster Monitor200 USER MANUAL 62330MN000 A01 15 of 51 Probe Sensor Figure 3 Probe Sensor Sidewall Sensor Figure 4 Sidewall Sensor Inline Sensor Figure 5 Inline Sensor...

Страница 16: ...his back plate Part 62310FA006 sits directly on the wall or cabinet surface and is secured by self tapping screws or industrial Velcro Sensor and control wiring access is at the back while access for...

Страница 17: ...nections to the Rooster Monitor200 alarm module Make any other optional connections to the alarm module 7 Align top of alarm module with tabs on back plate and secure in place with captive screw at bo...

Страница 18: ...a removable air screen which is washable and provides the sensor with a level of protection against tampering The Sidewall sensor should be located away from a sash opening or air moving device to avo...

Страница 19: ...330MN000 A01 19 of 51 Figure 11 Sidewall Sensor on Single Walled Cabinet Figure 12 Sidewall Sensor Mounting Single Walled Cabinet Self tapping mounting screw Extends through wall no ducting needed RJ1...

Страница 20: ...tallation method is required This method is also required for installations where the Sidewall sensor is mounted on the side but there are two sheet metal walls for the airflow to travel through In bo...

Страница 21: ...200 USER MANUAL 62330MN000 A01 21 of 51 Figure 15 Sidewall Sensor Mounting Dual Walled Cabinet Figure 16 Mounting Dimensions Wall 1 Front Figure 17 Mounting Dimensions Wall 2 Rear RJ11 connection to R...

Страница 22: ...tool access For this reason the RoosterTM can be ordered with an inline sensor which is quickly connected by splicing into the existing flow tube being used by another monitor The suggested installat...

Страница 23: ...ibration requires a Certifier passcode This is explained in the Access Tier Privileges section below The RoosterTM Monitor can be rebooted at any time by pushing the hard Home and Mute buttons simulta...

Страница 24: ...and Certifier functions you may set an identical passcode for both access tiers to make logging in and accessing key functions easier Calibration Calibration Airflow Sensor Once you have installed an...

Страница 25: ...er your Certify password If you have not set a custom password the factory default is 6425 Home Settings Calibration Calibrate 5 For your initial calibration keeping the default Normal value for Allow...

Страница 26: ...between m s and fpm units press the m s or fpm button on the top right corner of the keypad Press Enter to proceed and complete calibration Note For probe sensor ensure cap is off when establishing n...

Страница 27: ...reading 12 Rooster will use all measurement points to calculate and then display average face velocity for the sash high position establishing the low velocity set point Select Continue and follow th...

Страница 28: ...erify that the correct values have been entered and that there is a minimum 20 difference between them 3 Turbulence The RoosterTM Monitor200 has an Allowed Turbulence setting available to users who ar...

Страница 29: ...der to proceed you will need to enter your Certify password If you have not set a custom password the factory default is 6425 Home Settings Calibration Calibrate 4 Push the Calibrate Sash Position but...

Страница 30: ...oints is most common The alarm trip point is the air velocity value that will trigger an alarm The alarm clear point is the air velocity value that will clear an alarm state Once you have defined thes...

Страница 31: ...ck The RoosterTM Monitor200 supports Night Setback NSB for energy savings and greater HVAC system integration in building automation systems Users can toggle between three modes of operation detailed...

Страница 32: ...0 minutes if the unit is still in an active alarm state 60min A muted alarm will ringback audibly after 60 minutes if the unit is still in an active alarm state Alarm Delay To prevent nuisance alarms...

Страница 33: ...g indicates that an alarm state was triggered even if the device is no longer in an active alarm state This could occur when a user has stepped away from the workstation When latching is On a yellow b...

Страница 34: ...ess adjusts immediately as the setting is changed Pressing the Enter button will save the brightness setting and Cancel will leave the setting unchanged A timer is reset each time the screen is presse...

Страница 35: ...o 1 fpm This is the default setting 2 fpm Displayed units in FPM are rounded to 2 fpm 5 fpm Displayed units in FPM are rounded to 5 fpm 01 m s Displayed units in m s are rounded to 01 m s This is the...

Страница 36: ...A third units touch toggle of CFM is accessible on the main screen with volumetric flow enabled Air Changes Per Hour The Monitor200 displays velocity flow by default Airflow changes per hour ACH is a...

Страница 37: ...ormal troubleshooting steps follow the factory reset procedure to restore default settings and passwords BACnet Protocol Network Settings The MAC Address sets the main hardware address of the RoosterT...

Страница 38: ...Low Alarm Clear Analog Value 4 m s 0 to 40 99 R W Low air velocity value that will clear an alarm 85 High Alarm Trip Analog Value 5 m s 0 to 40 99 R W High air velocity value that will trigger an ala...

Страница 39: ...width 0 Valid if the selected duct area is rectangle Duct Dia Analog Value 15 cm 0 to 183 R W Airflow duct area diameter 0 Valid if the selected duct area is circle Face Velocity Analog Input 0 m s 0...

Страница 40: ...calibration Normal Setting Allowed Turbulence to High will help you overcome cabinet level airflow instability issues It should be reserved for rare circumstances where there is difficulty achieving...

Страница 41: ...1 R W Enable or disable the display of volumetric flow on the RoosterTM Disabled A non zero value for volumetric flow can be displayed on the RoosterTM provided duct information is entered Duct Shape...

Страница 42: ...tatus when not active Normally Open Relay3 Polarity Multi State Value 7 1 Normally Open 2 Normally Closed 3 Disabled 1 2 3 R W Relay out polarity status when not active Normally Open Relay1 Functional...

Страница 43: ...5 5 min 1 2 3 4 5 R W Amount of time alarm is paused on startup 30 sec Probe Type Multi State Value 15 1 Probe 2 Sidewall 3 Inline 1 2 3 R W Sensor type to be calibrated Sidewall Alarm Delay Multi Sta...

Страница 44: ...Character String Value 2 Standard serial number format R Serial number of RoosterTM RoosterTM information ID Data Rooster Caldate Character String Value 3 Standard date format R Calibration date NA Ro...

Страница 45: ...p on or off Fan is a separate on screen accessory button which can be used to turn a fan on or off Once the user links an accessory button to a relay output from the grid the button is enabled and dis...

Страница 46: ...b drive for future use Importing the EEPROM configuration file to other RoosterTM s in your facility ensures operational consistency and saves valuable setup time Plug a USB thumb drive into the Roost...

Страница 47: ...lts Thus the update should only be done during annual certification where the certifier would be recalibrating anyway OR if the client approves forcing a recalibration The first step of the firmware u...

Страница 48: ...The firmware release information listed next to firmware should have changed to the most current build If this is not the case repeat the upgrade procedure Power Cycle Procedure Procedure 1 Remove po...

Страница 49: ...ss the Mute key 2 The Enter Certify Password screen will display The factory default Certifier password is 6425 This can be changed by accessing the Advanced menu 3 The Factory Menu will display Depre...

Страница 50: ...Rooster Monitor200 USER MANUAL Rooster Monitor200 USER MANUAL 62330MN000 A01 50 of 51 GUI Map...

Страница 51: ...ured by DegreeC Such parts components or products may be warranted by third parties on a pass through basis The foregoing remedies shall not apply to any product failure caused in whole or in part by...