![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 970](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989970.webp)
Remote Commands
R&S
®
FPL1000
970
User Manual 1178.3370.02 ─ 09
Parameters:
<String>
String containing the description of the instrument.
Manual operation:
See
SYSTem:KLOCk
<State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State>
ON
LLO (local lockout)
OFF
GTL (go to local)
*RST:
OFF
Example:
SYST:KLOC ON
Activates LLO (remote control)
Manual operation:
See
SYSTem:LANGuage
<Language>
This function is used to emulate previous R&S signal and spectrum analyzers.
Parameters:
<Language>
"FSV" | "SCPI"
*RST:
SCPI
Example:
SYST:LANG FSV
The commands and settings for the R&S FSV instrument are
used.
SYSTem:LXI:INFO
This command queries the LXI settings.
Return values:
<LXIInfo>
<current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
•
<version>
•
<LXIClass>
•
<ComputerName>
•
<MACAddress>
•
<IPAddress>
•
<AutoMDIX>
Manual operation:
See
Configuring the R&S
FPL1000