Remote Commands
R&S
®
FPL1000
910
User Manual 1178.3370.02 ─ 09
Parameters:
<Distance>
Defines the distance that the limit line moves.
The unit depends on
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:STATe
<State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:THReshold
<Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S
FPL1000 uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<li>
Setting Basic Spectrum RF Measurement Parameters