Remote Commands
R&S
®
FPL1000
742
User Manual 1178.3370.02 ─ 09
//Defines a list with 3 entries and initiates the measurement with synchronization to the end.
//Analyzer produces a service request
//On service request:
SENS:LIST:POW:RES?
//Returns the results of the measurements, two for each frequency (peak and RMS power).
-----Initiliazing the measurement and querying results simultaneously-----
LIST:POW?
935.2MHZ,0,10,OFF,NORM,1MHZ,3MHZ,440us,0,
935.4MHZ,0,10,10,CFIL,30KHZ,100KHZ,440us,0,
935.6MHZ,0,10,20,CFIL,30KHZ,100KHZ,440us,0
//Defines a list with 3 entries, initiates the measurement and queries the results.
//Result example:
-28.3,-30.6,-38.1
11.6.16
Measuring the Pulse Power
All remote control commands specific to measuring the mean or peak pulse power
(e.g. bursts in various telecommunications standards) are described here. This mea-
surement is available via remote control only.
The Pulse Power measurement is a gated measurement that determines the power
over a particular number of pulses. The measurement is controlled by an external trig-
ger or the video signal. A separate trigger event is required for each burst included in
the measurement. In case of an external trigger source, the trigger level corresponds
to the TTL level. In case of a video signal, you can define any threshold.
The figure below shows the relations between the available trigger settings.
The measurement is always on trace 1, either with the peak detector to determine the
peak power or the RMS detector to determine the RMS power. Overall, you can config-
ure the measurement independent of the instrument setup with the commands listed
below only, which results in faster measurements.
The Pulse Power measurement is incompatible to other measurement functions (e.g.
marker functionality or statistics). If you use a command that controls those functions,
the R&S
FPL1000 aborts the Pulse Power measurement.
The R&S
FPL1000 also aborts the Pulse Power measurement if you end the remote
session.
Configuring and Performing Measurements