![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 716](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989716.webp)
Remote Commands
R&S
®
FPL1000
716
User Manual 1178.3370.02 ─ 09
See also
Suffix:
<n>
.
<m>
Return values:
<PeakPower>
Peak power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command queries the maximum positive peak time domain power. The query is
only possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<PeakPower>
Peak power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult?
This command queries the positive peak time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<PeakPower>
Peak power of the signal during the measurement time.
Usage:
Query only
Manual operation:
See
Configuring and Performing Measurements