Remote Commands
R&S
®
FPL1000
675
User Manual 1178.3370.02 ─ 09
<ri>
1..n
Selects the measurement range.
Parameters:
<RefLevel>
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Default unit: dBm
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME
<SweepTime>
This command defines the sweep time for a SEM range.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<SweepTime>
Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Default unit: s
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a SEM range on and
off.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation:
See
Configuring and Performing Measurements