Remote Commands
R&S
®
FPL1000
668
User Manual 1178.3370.02 ─ 09
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a SEM range on and off.
In case of high speed measurements, the state of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INSert
<Mode>
This command inserts a new SEM range and updates the range numbers accordingly.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
Parameters:
<Mode>
AFTer | BEFore
AFTer
Inserts a range after the selected range.
BEFore
Inserts a range before the selected range.
Manual operation:
See
"Insert before Range / Insert after Range"
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the measurement range.
<li>
1..n
Power class for which the limit is defined.
Configuring and Performing Measurements