![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 956](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987956.webp)
Remote Commands
R&S
®
FPL1000
956
User Manual 1178.3370.02 ─ 11
<FreeDiskSpace>
numeric value in bytes
Amount of free storage space on the R&S
FPL1000
<FileSize>
numeric value in bytes
Size of a single transducer file
<FileName>
string
Name of a single transducer file
Example:
SENSE:CORR:TRAN:CAT?
//Result: 2743,2312620544,720,'FactorGSM.TDF',2023,'FactorBTS.TDF'
Usage:
Query only
[SENSe:]CORRection:TRANsducer:COMMent
<Comment>
This command defines the comment for the selected transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Comment>
*RST:
(empty comment)
Manual operation:
See
[SENSe:]CORRection:TRANsducer:DATA
{<Frequency>, <Level>}...
This command configures transducer factors for specific trace points. A set of trans-
ducer factors defines an interpolated transducer line and can be stored on the instru-
ment.
Parameters:
<Frequency>
The unit for <Frequency> is Hz, which may or may not be omit-
ted. Frequencies have to be sorted in ascending order.
Default unit: Hz
<Level>
The unit for <Level> depends on
Example:
SENSe1:CORRection:TRANsducer:UNIT 'DB'
// Frequency Span 0 Hz to 4 Ghz
SENSe1:CORRection:TRANsducer:DATA 0,8,2GHz,5,4GHz,3
Creates the transducer points:
Manual operation:
See
Frequency
Level
0 Hz
8 dB
2 GHz
5 dB
4 GHz
3 dB
Configuring the R&S
FPL1000