![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 899](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987899.webp)
Remote Commands
R&S
®
FPL1000
899
User Manual 1178.3370.02 ─ 11
CALCulate<n>:LIMit<li>:LOWer:OFFSet
<Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
<li>
Parameters:
<Offset>
Numeric value.
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:SHIFt
<Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
<li>
Parameters:
<Distance>
Defines the distance that the limit line moves.
The unit depends on
on page 901.
Default unit: DB
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters