User Manual 1178.3370.02 ─ 11
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK2:FUNC:NOIS ON
Switches on noise measurement for marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:FUNC:NOIS:RES?
Outputs the noise result of marker 2.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
<State>
This command turns the noise measurement at the marker position on and off.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Manual operation:
See
See
"Switching All Noise Measurement Off"
10.8.8.10
Phase Noise Measurement Marker
The following commands control the phase noise measurement marker function.
Useful commands for phase noise markers described elsewhere
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:
PEAK]
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
Setting Basic Spectrum RF Measurement Parameters