![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 760](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987760.webp)
Remote Commands
R&S
®
FPL1000
760
User Manual 1178.3370.02 ─ 11
The EMI-specific filter types are available if the EMI (R&S
FPL1-K54) measurement
option is installed, even if EMI measurement is not active. For details see
Parameters:
<FilterType>
CFILter
Channel filters
NORMal
Gaussian filters
P5
5-pole filters
The 5-pole filter is not available for FFT sweeps.
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FPL1-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FPL1-K54) option
*RST:
NORMal
Example:
BAND:TYPE NORM
Example:
See
Chapter 10.6.17, "Programming Example: Performing a
Manual operation:
See
See
See
[SENSe:]BWIDth:VIDeo
<Bandwidth>
[SENSe:]BANDwidth:VIDeo
<Bandwidth>
This command defines the video bandwidth.
The command decouples the video bandwidth from the resolution bandwidths.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
AUTO is set to ON
Default unit: HZ
Example:
BAND:VID 10 kHz
Manual operation:
See
[SENSe:]BWIDth:VIDeo:AUTO
<State>
[SENSe:]BANDwidth:VIDeo:AUTO
<State>
This command couples and decouples the video bandwidth to the resolution band-
width.