Remote Commands
R&S
®
FPL1000
731
User Manual 1178.3370.02 ─ 11
<GateLength>
Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from
IMMediate
.
Range:
31.25 ns to 30 s
*RST:
0 s
Default unit: s
[SENSe:]LIST:POWer:STATe
<State>
This command turns the List Evaluation off.
Parameters:
<State>
OFF | 0
*RST:
0
10.6.15.2
Example: Performing List Evaluation
The following example shows a list evaluation with the following configuration.
No
Freq
[MHz]
Ref
Level
[dBm]
RF
Attenu-
ation
[dB]
El
Attenu-
ation
[dB]
Filter
RBW
VBW
Meas
Time
Trigger
Level
1
935.2
0
10
---
Normal
1 MHz
3 MHz
440 µs
0
2
935.4
0
10
10
Channel 30 kHz
100 kHz
440 µs
0
3
935.6
0
10
20
Channel 30 kHz
100 kHz
440 µs
0
-----Measurement with synchronization via service request-----
*ESE 1
*SRE 32
// Configures the status reporting system to produce a service request.
LIST:POW:SET ON,ON,OFF,EXT,POS,10us,434us
//Turns on the list evaluation, configures the global list evaluation settings and
//evaluates the peak and RMS power.
LIST:POW
935.2MHZ,0,10,OFF,NORM,1MHZ,3MHZ,440us,0,
935.4MHZ,0,10,10,CFIL,30KHZ,100KHZ,440us,0,
935.6MHZ,0,10,20,CFIL,30KHZ,100KHZ,440us,0;
*OPC
//Defines a list with 3 entries and initiates the measurement with synchronization to the end.
//Analyzer produces a service request
//On service request:
SENS:LIST:POW:RES?
//Returns the results of the measurements, two for each frequency (peak and RMS power).
-----Initiliazing the measurement and querying results simultaneously-----