Remote Commands
R&S
®
FPL1000
619
User Manual 1178.3370.02 ─ 11
CPOWer
Channel power measurements.
Returns the channel power. The unit of the return values
depends on the scaling of the y-axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
For SEM measurements, the return value is the channel power
of the reference range.
PPOWer
Peak power measurements.
Returns the peak power. The unit of the return values depends
on the scaling of the y-axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
For SEM measurements, the return value is the peak power of
the reference range.
Note that this result is only available if the power reference type
is set to peak power (see
OBANdwidth | OBWidth
Occupied bandwidth.
Returns the occupied bandwidth in Hz.
COBandwidth | COBWidth
<Centroid frequency>,<Frequency offset>
See
Chapter 6.2.6.2, "OBW Results"
Manual operation:
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:SELect
<MeasType>
This command selects a power measurement and turns the measurement on.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Parameters:
<MeasType>
ACPower | MCACpower
Adjacent channel leakage ratio (ACLR), also known as adjacent
channel power or multicarrier adjacent channel.
The R&S
FPL1000 performs the measurement on the trace
selected with
CPOWer
Channel power measurement with a single carrier.
The R&S
FPL1000 performs the measurement on the trace
selected with
Configuring and Performing Measurements