![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987050.webp)
Getting Started
R&S
®
FPL1000
50
User Manual 1178.3370.02 ─ 11
Figure 4-12: The "MultiView" tab
4.3.4
Performing Sequential Measurements
Although only one measurement can be performed at any one time, the measurements
configured in the active channel setups can be performed sequentially, that means:
one after the other, automatically, either once or continuously.
1. Tap the "Sequencer" icon in the toolbar.
2. Toggle the "Sequencer" softkey in the "Sequencer" menu to "On".
A continuous sequence is started, i.e. each channel setup measurement is per-
formed one after the other until the Sequencer is stopped.
Trying Out the Instrument